Simple operation process of sulfate reagent

Author: 同奥科技
Abstract: 本文提供了水样硫酸盐浓度测试的简易操作流程。首先,对水样的硫酸盐浓度进行预判,若浓度在0-200mg/L范围内可直接测试,若超出需稀释至该范围后再测试。操作过程中需确保硫酸盐试剂充分摇匀,并注意超量程提示。测试时,若水样显示柠檬黄色,颜色深浅与浓度相关。测试步骤包括稀释、上机操作,并强调注意事项,如避免交叉污染、选择合适的移液用具及定期校准移液枪等,以确保测试结果准确可靠。

Concentration prediction:Before testing the water sample, the sulfate concentration value of the water sample should be predicted, and the excess concentration should be diluted and measured0-200mg/L, you can directly follow the operation steps; If the sulfate concentration is expected to exceed 200mg/L, the water sample should be diluted to the range of 0-200mg/LThe assay is performed.

Notes:Sulfate reagent 2 must be shaken well before use.

1. When the predicted value is in the range of 0 - 200mg/L

In the presence of sulfate, the water sample shows lemon yellow, and the darker the color, the higher the concentration value of the water sample. Analytical Procedure: Sulfate; Watch out for overrange tips!

Sulfate reagent simple operation process with Figure 1

2. When the predicted value exceeds 200 mg/L

The water sample needs to be diluted to the above 0-200 mg/L concentration value, and then the corresponding range is manipulated, and the measurement result is multiplied by the dilution factor.

3. Hands-on test steps

Sulfate reagent simple operation process with Figure 2

4. Precautions

◆ If there is sulfate, the water sample shows lemon yellow, and the darker the color, the higher the concentration value of the water sample.

◆ Pay attention to cross-contamination, use different measuring utensils for different liquids (pay attention to changing the tip when using a pipette).

◆ Use appropriate pipetting tools to take liquids, and large range measuring tools cannot measure small volumes of water samples, such as: 5mL pipettes cannot be used to remove 2mL and below liquidsFor liquids less than or equal to 2mL, it is recommended to pipette twice with a 1mL pipette.

◆ When using a pipette, it is necessary to calibrate the pipette frequently (it is recommended to do so once a week), and pay attention to replacing the tip during use.