Huitai GHP Incubator Instruction Manual

Installation and commissioning

1. Install

The water-proof electric heating Incubator should be placed in a room with good ventilation and no strong vibration. Inflammable and explosive temperature of the installation room is 5°C-40°C, the relative humidity is ≤85%, and it is non-corrosive. gas.

2. Debugging

1. Close the inner and outer glass doors of the Incubator.

2. Plug in the power plug and turn on the power switch. At this time, the low water level indicator light will give an alarm. Connect one end of the hose with an inner diameter of 4 minutes to the water inlet and the other end to the water source (faucet). Add water until the low water level alarm light is off, and continue to add water (note: add water slowly) until the high water level alarm or the high water level alarm light is on. (Note: The high water level alarm will automatically stop after 5 seconds, and the alarm indicator light will not go out. It will not affect the heating).

3. Set the temperature.

4. Observe whether the temperature rises. If the temperature rises and keeps constant, it means that the equipment is operating normally.

5. Turn off the power.


temperature setting

1. Click the setting button to enter the temperature setting state, modify the required value through the increase, decrease and shift keys; then press the setting button to enter the time setting state, through the increase, decrease and shift key to modify the required value, and then press the set key to save and exit the setting state.

2. When the time is set to "0", it means that there is no timing function. When the set time is not "0", the timer up , the buzzer beeps for 30 seconds, press and hold the down key for 4 seconds, the program will restart , and the buzzer can be silenced by pressing the up key.

3. Temperature internal parameters

Press and hold the set button for about 3 seconds, and the password prompt "LC" will appear. You can change the password to "3" through the increase, decrease and shift keys. Click the set button to enter the temperature internal parameter state. Click the set button to Modify each parameter, and then press and hold the set button for 3 seconds to exit this state, and the parameters are automatically saved. If no key is pressed within 30 seconds, the state is automatically exited, and the current parameters are not saved.

Huitai GHP Water-Jacketed Incubator instruction manual with picture 1

Safety Precautions

1. The cabinet must be effectively grounded to ensure safety.

2. Box items cannot be placed too densely

3. Do not wipe the distribution line with a wet cloth

4. This type of non-explosion-proof product, flammable and explosive items cannot be put into the box for high-temperature baking test.

5. If the fuse is damaged due to a fault, a fuse of the same specification should be replaced when repairing.


maintenance and maintenance

1. The Incubator should always be kept clean, and the door glass should be wiped with a soft cotton cloth. Do not wipe with a reactive chemical solution to avoid chemical reactions and scratches on the glass.

2. If the Incubator is not used for a long time, it should be covered with a plastic protective bag and placed in a dry room to avoid the use .