How to use the rigid foam Water Absorption Tester

Test Methods

▲The size of the measured object: length 150MM, width 150mm, thickness 75mm; volume not greater than 500cm3;

▲Ambient temperature: 23±2℃ Ambient humidity: 50±5﹪

▲Soaking liquid: distilled water that is placed for 48 hours after distillation;

▲ Cut the rigid foam into a square test object with a length of 150mm, a width of 150mm, and a thickness of 75mm.

▲Put the sample on the balance, weigh the dried sample weight (m1), accurate to 0.1g, and record it.

▲Immerse the hanging cage in distilled water, remove the air bubbles on the surface of the hanging cage, hang it on the lower hook of the balance, weigh its apparent weight (m2), accurate to 0.1g, and record it. After remembering, remove the hanging cage .

▲ Loosen the two fixing screws on the sleeves on the two sides of the hanging net cage respectively, then place the sample on the disc and center it; lower the pressure rod to press the sample, fasten the two ends of the pressure rod to the sample, and rotate Tighten the two setscrews on the 2-sided sleeve.

▲ Re-immerse the hanging cage with the sample installed in distilled water, and use a soft brush and stir to remove the air bubbles on the surface of the hanging cage and the hard foam.

▲ After soaking for 96 hours, hang the hanging net cage containing the sample on the weighing hook under the balance, weigh its apparent mass (m3), accurate to 0.1g, and remember it.

System composition

Rigid foam water absorption test system is composed of: rigid foam water absorption test calculation software, special cage for sample loading, test sample Water Tank, rigid foam quality measuring instrument, special sample slicer, glass slide, etc. *** Partial composition.

Rigid Foam Water Absorption Tester Slicer

①Instructions for use

The Rigid Foam Specimen Slicer is part of Rigid Foam Water Absorption.

The main structure of the rigid foam sample slicer includes a sample placement slot, a sample thickness regulator, and a sample tightening screw. The function of the sample placement regulator is to place the sample and adjust the thickness of the sample slice.

②Technical parameters:

●Slicer minimum slice thickness: 0.01-0.04mm

●Sample size: 50*50*100mm

●Dimensions: 150*150*130mm

●Effective stroke: 25mm

●Overall weight: 11KG

③How to use:

●Adjust the thickness regulator of the test piece to the 25mm scale, loosen the sample tightening screw, put the sample with a size of 50*50 into the placement groove, and tighten the sample clamping screw to compress the test piece.

●Use a slicing knife to cut off the excess sample along the surface of the sample cutter platform. In order to ensure that the upper part of the sample is flat, you can adjust the thickness of the slice to 0.3mm, and use the slicing knife to align the sample.

●The slice thickness regulator adjusts the required slice thickness (clockwise for extruded samples, counterclockwise for samples returned to the sample slot), but the thickness of the blade should be added, the thickness of the slice blade is about 0.15-0.2mm, for example: cutting When the thickness is 0.1mm, it should be screwed out by about 0.3mm, and then sliced ​​with a slicer.

 