After the Leeb Hardness Tester has been used for a period of time, due to improper use or other problems that will affect the accuracy of the instrument, it needs to be recalibrated and maintained. This article mainly gives a brief introduction to the calibration and maintenance of the Leeb Hardness Tester.
After a period of use, especially after testing some relatively rough materials, the dots of the impact body may be a little worn, which will inevitably lead to a more or less affected measurement accuracy. In this case, the instrument needs to be recalibrated as follows:

Note: If the error of the instrument is within the basic range, in order not to affect the long-term stability of the instrument, please do not calibrate it lightly.
1) After 1000-2000 times of use, clean the conduit and impact body of the impact device with a nylon brush. When cleaning the conduit, first unscrew the support ring, then take out the impact body, and screw the nylon brush into the tube counterclockwise. Pull it out after reaching the end, repeat this 5 times, and then install the impact body and support ring;
2) After use, the impact body should be released;
3) It is strictly forbidden to use various lubricants in the impact device.
4) When the standard Rockwell hardness block is used for verification, if the error is greater than 2HRC, it may be that the ball head is worn out, and replacement of the ball head or impact body should be considered.
The above is about the calibration and maintenance of the Leeb Hardness Tester. For details, please refer to the "Leeb Hardness Tester Instruction Manual"