How does a Rotational Viscometer reduce measurement errors?

Abstract: 本文介绍了降低旋转粘度计测量误差的有效方法。为了确保准确的粘度测量,建议采取以下措施:使用恒温水槽控制液体温度,并保持一致;在测量前清洁转子和保护框,并安装保护框;将转子置于容器中心并使其液面标志与液面平齐,保持主机水平;让转子与液体温度一致后再开始测量;从高速测量改为低速测量时,应先复位电机,然后再设置低转速;根据液体粘度选择合适的转子,并在低速测量高粘度液体时延长测量时间以稳定数据;定期检查和调整主机水平状态;使用标准粘度液进行定期性能检测。

The Rotational Viscometer is used to measure the viscous resistance and dynamic viscosity of various liquids, and is widely used in the viscosity measurement of various fluids such as oil, paint, food, paper, cosmetics, chemicals, coatings, and capsules.Although the Rotational Viscometer is simple in structure and easy to use, if it is not used properly, it will not be able to obtain accurate measurement results and affect product quality. There are many factors that will cause measurement errors during the test, how to avoid these errors?

  • Use a constant temperature Water Tank to control the temperature of the liquid to be measured and make each measurement under the same temperature adjustment.

  • Do a good job of cleaning the rotor and protective frame before measurement. When using L1~L4, R1~R7 rotors for measurement, the protective frame needs to be installed, otherwise the accuracy of the measurement data will be affected.

  • When measuring, the rotor should be placed in the center of the container, and the rotor liquid level mark should be flush with the liquid level, and the main engine should be in a horizontal state.

How does a Rotational Viscometer reduce measurement errors?  Picture 1

  • When the rotor is immersed in the liquid to be measured, it should wait for ten minutes to make the temperature of the two consistent before taking the measurement.

  • When changing from high-speed measurement to low-speed measurement, you should first press the <Reset> button to stop the motor, then set it to a low speed and wait for the liquid level to stand still and then press the <Measurement> button to measure, so as to eliminate the rotation inertia of the liquid. Measurement error.

  • When measuring low-viscosity liquids, first select the L1 rotor, and when measuring high-viscosity liquids, first select the L4 rotor, and then decide whether to change the rotor and adjust the speed according to the displayed torque value.

  • When measuring high-viscosity liquids at low speed, it is necessary to prolong the measurement time to stabilize the data before reading the results.

How does a Rotational Viscometer reduce measurement errors?  Picture 2

  • During the measurement process, due to lifting and lowering the main engine, changing the rotor or the liquid to be measured, the main engine will often lose its level state, and the level should be checked and readjusted in time.

  • Regularly or irregularly inspect the main engine with standard viscosity liquid to judge and check its performance and working condition. (Standard viscosity liquid with corresponding parameters should be selected according to the commonly used viscosity measurement range.)