Maintenance and Maintenance of Dissolved Oxygen Electrode

Abstract: 本文介绍了溶解氧电极的维护与保养方法,重点强调了日常使用中的清洁和保护措施。溶解氧电极由电极棒、电极膜帽及填充液组成。为了保持其性能,应确保电极膜帽洁净,避免接触尖锐物质。当电极棒的阴极与阳极发黑时,需用抛光布轻轻擦拭,直至亮色恢复,并用酒精和蒸馏水进行清洗。如果长时间不使用电极,应卸下膜帽,倒出填充液,并清洗电极组件,最后将电极避光存放。

Dissolved oxygen electrode is a sensor used to measure the concentration and content of dissolved oxygen in water quality. A complete polarographic dissolved oxygen electrode is composed of electrode rod, electrode membrane and filling liquid. In daily use, the membrane of the dissolved oxygen electrode should always be kept clean and away from sharp objects. When the cathode and anode of the electrode rod appear black, you need to use a polishing cloth to gently wipe the surface until a bright color appears. After wiping, the cathode and anode can be rinsed with alcohol, and then cleaned with distilled water.

Maintenance and Maintenance of Dissolved Oxygen Electrode Figure 1

If you do not use the dissolved oxygen sensor for a long time, please remove the membrane at the front of the electrode, pour out the filling solution and rinse the membrane with distilled water, the anode and cathode of the electrode. Use a clean cloth to wipe off any remaining water droplets on any part of the electrode, then install the membrane cap to the electrode rod and store it away from light.

Products required