Back
Knowledge

Fangrui surface tensIon Meter test data deviation, how to solve?

Abstract: 该文章介绍了如何解决方瑞表面张力仪测试数据偏差大的问题。首先,建议彻底清洗铂金片,使用流水冲刷并用纯酒精灯烧红以去除杂质。再有就是应清洗玻璃器皿并用样品润洗,确保样品纯净。第三,进行砝码校正时需按照说明书操作,避免额外动作,以确保准确校准。最后,建议使用瓶装饮用蒸馏水进行测试,以避免因水质问题导致的数据偏差。如果问题仍未解决,可咨询南北潮技术。

Surface tensiometer

During the experiment, the Surface Tensiometer test data deviates greatly, what should I do? How can I manually check the correction myself?

The Surface Tensiometer is a very precise instrument and equipment, and the test must be very meticulous. A small place will have a great impact on the data. When encountering a large data deviation, you can first check and confirm three points:

The first point is that the platinum sheet should be cleaned. The thin lens should be rinsed with running water first. After rinsing, it should be burnt with an alcohol lamp and put on the alcohol lamp. Burn the entire lens red. Remember that it is red, and put it on top. If it does not move for 20 to 30 seconds, it will burn the alcohol and burn the entire platinum plate red. If the alcohol burned is pure alcohol, it cannot be impure alcohol. Pure alcohol is this kind of pure alcohol with a concentration of more than 99.5%. Alcohol burns.

The second point is to wash the glassware, rinse it with running water, and then rinse it with the sample to be tested. Pour off the liquid after moistening the snow, and then pour the tested sample into it.

The third point is to calibrate the weight of the machine. There are steps for weight calibration in the instruction manual. Pay attention when calibrating. You can’t do other movements that are redundant. After hanging it up, wait for about 20 seconds and there will be a beep. After the sound is over, it will have a 600.0. Then take the weight down. At this time The number slowly returns to zero, which means the calibration is complete.

The last point, the measured water, it is recommended to go outside to get that kind of bottled distilled water for drinking, such as drinking distilled water, a bottle is three to five hundred milliliters, if this kind of water, its source is no problem, guaranteed, so I say this It is very convenient to know if the water is planted, because there is still no problem with the sample, and sometimes the customer uses other water to cause data deviation.

The above is for reference. If you still can’t solve the problem, you can consult the North and South Tide Technology.

Products required
Relevant knowledge
Guess you like it