Constant temperature and humidity Test Chambers are commonly referred to as high-low temperature alternating damp heat Test Chambers, high-low Temperature Test Chambers, high-low temperature damp heat Test Chambers, etc. It is generally applicable to the quality inspection of mobile phones, electronics, electrical appliances, communications, instruments, vehicles, plastic products, hardware, food, chemicals, medical care, building materials, aerospace and other products.
The constant temperature and humidity Test Chamber is a high-end precision equipment, and the damage of one of the components may cause the Test Chamber to fail to start and operate normally.
Troubleshooting and solutions when the constant temperature and Humidity Chamber cannot start normally:
1. Whether the connection of the constant temperature and Humidity Chamber is in good contact: reconnect;
2. Whether the main switch contacts of the constant temperature and Humidity Chamber are in good condition/burned or damaged: repair/replace;
3. Whether the contactor or relay in the constant temperature and Humidity Chamber is in good contact/burned or damaged; whether the over-temperature setting of the instrument is correct/burned or damaged: reset/replace;
4. Whether the phase sequence connection of the constant temperature and Humidity Chamber power line (phase sequence relay) is correct/burned or damaged: reconnect/replace;
5. Whether the pressure controller of the constant temperature and Humidity Chamber is adjusted correctly/burned and damaged: readjust/replace;
6. Whether the non-heating connection of the constant temperature and Humidity Chamber is in good contact: reconnect;
7. Whether the instrument display in the constant temperature and humidity room is normal / whether there is output: maintenance / replacement;
8. Whether the instrument settings in the constant temperature and Humidity Chamber are correct: reset;
9. Whether the shot blasting switch in the constant temperature and Humidity Chamber is turned on/whether it is burned or damaged/whether the contactor is connected or burned: re-open/replace;
10. Whether the heating switch of the constant temperature and Humidity Chamber is turned on / whether the contact is good / whether it is burnt out: turn on the switch / repair / replace;
11. Whether the actuators (contactors, relays, thyristors, solid state relays) in the constant temperature and Humidity Chamber have output: replace;
12. Whether the actuators (heating tubes, heating wires) of the constant temperature and Humidity Chamber are burned or damaged: replace;
13. Whether the controller, regulator, and electric contact thermometer of the constant temperature and Humidity Chamber are in good condition/burned or damaged: repair/replace;
14. Whether the heating of the constant temperature and Humidity Chamber is not controlled by the instrument display screen/whether the control is normal: maintenance/replacement;
15. Whether the instrument settings in the constant temperature and Humidity Chamber are correct/whether it is manual output: reset;
16. Whether the actuator contactor and relay contacts in the constant temperature and Humidity Chamber are adhered or burned / whether the thyristor and solid state relay are faulty: replace;
17. Whether the heating tube and heating wire of the actuator in the constant temperature and Humidity Chamber are grounded: repair/replace.