What should be paid attention to when using a constant temperature and humidity Test Chamber?

Abstract: 文章详细说明了使用恒温恒湿试验箱时需要注意的事项。观察箱内变化时,可开启室内灯,但在低温运转(0℃以下)时应避免频繁开门,以防内部结冰。低温操作完成后,应进行干燥处理,或设定常温条件,以防止结冰和玻璃结雾。定期保养冷凝器、排除加湿桶入水管的空气,确保设备正常运作。操作中应避免打开箱门,并确保设备安全接地。避免短时间内频繁启动冰冻机组,并使用外部电源控制发热试料,确保试验安全。
  • If you want to observe the changes in the box during the test, you can turn on the switch of the indoor light (LIGHT), and know the changes inside through the window.

  • If the constant temperature and humidity Test Chamber is operated below 0°C, try to avoid opening the door, because at low temperature, if the door is opened, it will easily cause the internal evaporator and other parts to freeze, especially the lower the temperature, the more serious the situation , if it needs to be opened, the opening time should be shortened as much as possible.

  • When the low-temperature operation is completed, be sure to set the temperature condition at 60°C and carry out drying treatment for about half an hour, so as not to affect the measurement time or freezing phenomenon of the next operating condition.

  • The radiator (condenser) of the freezer should be maintained regularly and kept clean.

  • When the humidifying barrel enters the water pipe, the trapped air needs to be completely discharged to prevent water from entering.

*Special attention: After the machine has finished running at low temperature, if there is no work in the fifth point above, it is also necessary to set the normal temperature condition (30±5°C) so that the temperature in the box is at normal temperature when it is shut down, so as not to affect the temperature in the box. icing and window glass fogging.

  • During operation, unless absolutely necessary, please do not open the door, otherwise it may lead to the following adverse consequences: 

- High temperature and humidity rushing out of the box...very dangerous

- The inside of the door is still hot... causing damage

- High temperature air may trigger fire alarm and cause false operation

  • Please note that the machine needs to be safely grounded to avoid electrostatic induction

  • Avoid turning off and on the freezer within three minutes

  • If a heating sample is placed in the box, please use an external power supply for the control of the sample power supply, do not use the local power supply directly

  • Circuit breaker, temperature over-temperature protector, provide the test product of this machine to protect the safety of the operator, so please check it regularly

  • Correctly install the gauze of the wet bulb to ensure the correct relative humidity

  • It is absolutely forbidden to test explosive, flammable and highly corrosive substances