1. Try to avoid frequent start and stop of the thermal shock test room (try not to restart the operation within 5 minutes after the stop of the thermal shock test room), so as not to increase the mechanical load of the refrigeration system compressor and affect the service life of the unit.
2. It is strictly forbidden to test flammable, explosive, volatile, and corrosive substances in the thermal shock Test Chamber indoors, otherwise it will cause accidents, damage the machine, and invalidate the test results.
3. The placement of the test object should not affect the airflow balance and the test equipment in the environment. Otherwise, it will affect the performance of the thermal shock Test Chamber and the accuracy of the test. On any cross-section perpendicular to the wind direction of the internal circulating airflow, the total area of the test object shall not be greater than 1/3 of the cross-sectional area of the working chamber.
4. The door of the test room must be closed tightly before operation, otherwise it will cause gas leakage in the work room, and the test does not meet the performance requirements.
5. During the operation of the thermal shock Test Chamber , please do not touch and check the system components of the machine with your hands to avoid accidents such as electric shock or motor impact. So please stop the machine and turn off the power before repairing.
6. If the thermal shock Test Chamber operates below 0 ℃ during the low temperature test, try to avoid opening the Test Chamber door, because opening the Test Chamber door during the low temperature test will easily cause frosting on the evaporator and other components of the refrigeration system. The lower the test temperature, the more serious the effect. If it is necessary to open the door, the opening time should be shortened as much as possible.
7. After using the thermal shock Test Chamber to complete the low temperature test operation, the test temperature condition should be set to 30 ℃, and dry for about half an hour (return to normal temperature, the thermostat can be set to automatic function), then close and open the test box. After the temperature returns to room temperature, the test products can be picked and placed so as not to affect the test performance.
8. When the thermal shock Test Chamber is operating under high temperature and high humidity conditions, it is strictly forbidden to open the door, otherwise it may cause the following adverse consequences:
(1) The high-temperature and high-humidity gas rushes out of the high-Temperature Test Chamber sharply, which may easily cause burns;
(2) The inside of the door, the test area and the surface of the sample are still at high temperature, which is easy to burn your hands;
(3) High temperature air may trigger a fire alarm and cause malfunction.