List Magnetik MEGA-CHEK Pocket FE platings Thickness Gauge

SE
German List Magnetik MEGA-CHEK Pocket FE coating platings Thickness Gauge Measurement principle magnetic Induction, Measurement range 0~5000μm, resolution 0.1μm | 1μm | 0.01mm, Pocket FE coating platings Thickness Gauge applied to all nonmagnetic coating platings on steel thickness measurement.
$1563.00/set
PromotionLoading...
Coupon

List-Magnetik MEGA-CHEK Pocket FE Coated platings Thickness GaugeSPEC

List-Magnetik MEGA-CHEK Pocket FE Coated platings Thickness GaugeDetails

List-Magnetik MEGA-CHEK Pocket FE Coated platings Thickness GaugePacking list

SKU
NB011678
Measurement principle
magnetic Induction
Measurement range
0~5000μm
margin of error
±1μm/±1~5%
resolution
0.1μm|1μm|0.01mm
Minimum measurement surface
Φ4mm
Minimum radius of curvature
Convex 4mm Concave 38mm
probe
PF-5
Calibrated mode
One point calibrated, two point calibrated
Unit display
μm/mils
statistical function
Do not have
icon function
none
Interface
Do not have
Data storage
Do not have
Power Supply
1.5V AA
Standards
ISO2178 and ISO2360
system language
English, German optional
Display screen
3 inch LED display
Weight
137g
Dimension
105*65*26mm
MEGA-CHEK Pocket FE
Coated platings Thickness Gauge
List Magnetik MEGA-CHEK Pocket FE platings Thickness Gauge Picture

The MEGA-CHEK coating Thickness Gauge uses a high-precision digital measurement probe in which the analog signal in the probe can be digitized and transmitted directly to the host computer. This technique avoids interference and maximizes the accuracy of repeated measurements. The high-precision probe is connected to the main machine through a special cable, and the probe with different functions only needs to be replaced, without changing the cable, which is extremely cost-effective.

The operation is very simple, one-click operation.

Equipped with ASR technology (automatic statistics function)

1. Instrument selection:

MEGA-CHEK Pocket FE Type: Equipped with a ISO2178-compliant magnetic induction probe. It is applied to the measurement of the thickness of all non-magnetic coatings on steel, such as paints, plastics, enamels, chromium, zinc, etc.

MEGA-CHEK Pocket FN Model: Equipped with a combination of magnetic induction and eddy current probes for ISO2178 and ISO2360. It is used to measure the thickness of all insulating coatings on non-ferrous metals (aluminum, zinc, copper, etc.), such as anodized films, paints, coatings, etc. It is also used to measure the thickness of all non-magnetic coatings on steel, such as paints, plastics, enamels, chromium, zinc, etc.

2. Instrument parameters 

Model

MEGA-CHEK  Pocket FE

MEGA-CHEK  Pocket FN

structure

Split Machine (Fe)

Split machine (Fe/NFe)

Meets standards

ISO 2178, ISO 2360,

use

It is applied to the measurement of the thickness of all non-magnetic coatings on steel, such as paints, plastics, enamels, chromium, zinc, etc.

It is used to measure the thickness of all insulating coatings on non-ferrous metals (aluminum, zinc, copper, etc.), such as anodized films, paints, coatings, etc.

It is also used to measure the thickness of all non-magnetic coatings on steel, such as paints, plastics, enamels, chromium, zinc, etc.

Measuring range:

FE:  0-5000um

FE:0-5000um

NFE:0-2500um

Display Resolution:

1-  100um: 0.1um

> 100um: 1 um

   >2000um: 0.01 mm

Accuracy:

<  100um:  ±1um

  100 - 1000um:  ±1%

  1000 - 2000um: ±3%

>2000um:  ±5%

Available Probes:

PF-5

PFN-52D

Minimum Area:

Φ4mm

Φ6mm

Minimum radius of curvature convex:

4mm

FE:4mm

NFE:6mm

Minimum radius of curvature concave:

38mm

38mm

Power Supply:

2X1.5V AA BATTERY

Calibration:

It has a single/two-point correction function

Shutdown:

Manual/automatic shutdown

Unit Conversion

Metric to Imperial Conversion Function, UM/MILS:

Statistics

Maximum, Minimum, Mean, Standard Deviation, Relative Standard Deviation (ASR Technique)

Display:

3  ? inch LED display

Size:

105  X 65 X 26 mm

weight

137g

Device configuration

Split main unit, PF-5 standard magnetic induction probe, substrate test plate (Fe), standard calibrator (2), operation manual, battery, packaging box.

Split main unit, PFN-52D magnetic induction/eddy current dual-function probe, matrix test board (Fe/NFe) with one each, standard calibrator (two), operation manual, battery, and packaging box.