China

Zibo Coulomb KLK301 open flash point automatic measuring instrument, measuring range 40~400 ℃

SE
Adopt ARM technology, equipped with a large-screen touch screen, support automatic heating, ignition, flash point locking and printing, in line with national standards, suitable for petroleum, chemical and other fields, temperature range 40~400 ℃
$2838.00/set
PromotionLoading...
Coupon
Model:

zibo KLK301 Open flash point automatic measuring instrumentSPEC

zibo KLK301 Open flash point automatic measuring instrumentDetails

zibo KLK301 Open flash point automatic measuring instrumentPacking list

SKU
NB012156
display
Color liquid crystal display
Temperature range
40~400℃
Detection type
Opening flash point or ignition point
Temperature Sensors
Platinum resistance
Sexual repeability
不大于8℃ (GB/T3536);≤150℃  4℃(GB/T267);>150℃  6℃(GB/T267)
Ignition method
Electronic ignition
Information storage
1000 measurement results can be stored
Chill down
forced air cooling
Printing method
Dot matrix thermal printer
Self-test function
Lifting rods, swipe rods, printing, etc
Power
AC 220V ± 11V; frequency 50Hz ± 2.5Hz
power
≤600VA
Environment temperature
10~35℃
Relative Humidity
≤90%
Weight
About 17kg
KLK301
Open flash point automatic measuring instrument
Adopt ARM technology, equipped with a large-screen touch screen, support automatic heating, ignition, flash point locking and printing, in line with national standards, suitable for petroleum, chemical and other fields, temperature range 40~400 ℃
Zibo Coulomb KLK301 open flash point automatic measuring instrument, measuring range 40~400 ℃ Picture

Introduction

KLK301 open flash point automatic Tester is a new type of flash point automatic Tester designed with ARM microprocessor technology. The instrument adopts a large-screen color LCD display, Chinese menu touch screen control, and a complete electric ignition method. It has the functions of automatic heating, automatic scanning, automatic ignition, timely selection of standard programs, automatic temperature tracking and display, automatic locking of flash point value and automatic printing of results, atmospheric pressure correction and forced air cooling.

Zibo Coulomb KLK301 open flash point automatic Tester has reasonable structure, simple operation, stable and reliable measurement results. It is a satisfactory instrument for open flash point determination.

application standard

KLK301 complies with the standards GB/T3536-2008 and GB/T267-88.