Stone light meter JFL-B60 (ordinary type) JFL paints, Stone and other gloss Measurement Instruments

SE
JFL JFL-B60 (common type) Stone light meter is a Gloss Meter with a Measurement angle of 60 °. JFL-B60 Stone light meter is used for the measurement of paints, coatings, Stone, flooring and other surface Glossiness.
$234.00/set
PromotionLoading...
Coupon
Certificate

JFL JFL-B60 (normal type) GlossmeterSPEC

JFL JFL-B60 (normal type) GlossmeterDetails

JFL JFL-B60 (normal type) GlossmeterPacking list

SKU
NB001292
Geometry
60°
Measurement range
0~199.9GU
Measuring Areas
10*20mm
Error of indication
±1.2GU
Standards
ISO-2813、ASTM-D2457、GB9754、GB9966、GB8807
Operating language
Chinese
Dimensions
114*35*65mm
Weight
300g
Power
One AA battery
Stability
±0.4GU
Measurement port size
14*28mm
Working Voltage
1.5V
Application
Paint, Coating, Printing ink, Plastic, Stone, Flooring, Wood Furniture Surface, etc
Measurement port shape
ellipse
JFL-B60 (normal type)
Glossmeter
Stone light meter JFL-B60 (ordinary type) JFL paints, Stone and other gloss Measurement Instruments Picture

Introduction

JFL JFL-B60 (ordinary type) stone light meter is a general-purpose gloss meter. It is a gloss test instrument produced according to the national inspection regulation JJG696 standard. It has good measurement stability and high precision. It is a small and compact stone light meter. With the automatic zero calibration function, only one AA battery can be used to start measuring the surface gloss of the object. The projection angle of JFL-B60 (ordinary type) gloss meter is 60 degrees, the measurement display range is 0.0~199.9Gs, and the indication error is less than 1.2.

Applications

JFL-B60 (Common Type) Stone Light Meter is suitable for measuring paint, paint, ink, plastic, stone, floor, wooden furniture and other surfaces.

product operation

1. Press the switch of the JFL-B60 stone light meter, place the gloss meter on the black glass calibration plate, and turn the adjustment knob to make the displayed value the same as the nominal value of the standard plate.

2. Put the gloss meter JFL-B60 on the white ceramic standard plate, the error between the displayed value and the nominal value of the white ceramic standard plate should not be greater than ±1.2 gloss units, otherwise, the standard plate is dirty and other problems should be considered.

3. When the error is out of tolerance, calibrate again according to the previous two steps.

4. Measure the sample after calibration, when the measured value exceeds 199.9 gloss units, the display overflows "1".

product structure

JFL-B60 stone light meter is composed of calibration knob, display window, switch key, knob type battery cover, measurement positioning mark, measurement caliber and other parts.

JFL JFL-B60 (common type) stone photometer structure diagram