JFL JFL-BZ60 (intelligent) gloss meter is an intelligent gloss meter with a projection angle of 60 degrees. Different from the ordinary type, the intelligent JFL-BZ60 gloss meter uses four small Ni-MH rechargeable batteries and can display The range is 50% higher than that of the JFL-B60 gloss meter, and there is no need to calibrate the nominal value. JFL-BZ60 (intelligent) gloss meter has a measuring range of 0.0~300.0Gs, the measurement window is elliptical, and the size is 14x28mm. It is a general-purpose gloss meter.
1. Press the switch of the JFL-BZ60 (intelligent) gloss meter, and the gloss meter is in the measurement mode.
2. Put the JFL-BZ60 intelligent gloss meter on the sample, and the displayed value is the gloss value of the sample.
3. When the measured value is greater than 300.0 gloss units, the word "OVER" will be displayed.
1. Put the measurement window of the JFL-BZ60 intelligent gloss meter on the black velvet board, press the "set number/zero" button, and the displayed value is 0.0.
2. Use the black glass standard plate instead of the black velvet class, put it under the measurement window of the JFL-BZ60 gloss meter, press the "calibration/measurement" button, enter the setting and loading, and then display "000.0" and the first "0" flashing.
3. Press the "Shift" key to move the flickering digit to the right, and press the "Number/Zero" button to change the value of the "blinking" digit from 0-9-0. By pressing the "Shift" and "Number/Set" "Zero" key causes the displayed value to be the same as
The nominal value of the black glass standard plate is the same, which means that the setting by pressing the "Calibration/Measurement" button is completed .
4. Put the JFL-BZ60 intelligent gloss meter measurement window on the white ceramic standard plate, the error should not be greater than ±1.2 gloss units, otherwise the standard plate should be considered dirty, etc. question.
5. After the calibration is completed, the glossiness of the sample can be measured.