China

Metal Surface Glossiness Detection Instrument JFL-B60M JFL 60 Degree Glossiness Detection Instrument

SE
The JFL-B60M Metal Surface Glossiness Detection Instrument is a Glossiness Detection Instrument with a Measurement angle of 60 degrees, and the JFL-B60M Glossiness Measurement Device is used for the measurement of metal surface Glossiness.
$285.00/set
PromotionLoading...
Coupon
Certificate

JFL JFL-B60M (common type for metal) GlossmeterSPEC

JFL JFL-B60M (common type for metal) GlossmeterDetails

JFL JFL-B60M (common type for metal) GlossmeterPacking list

SKU
NB001298
Geometry
60°
Measurement range
0~1999GU
Measuring Areas
10*20mm
Error of indication
±2GU
Division Value
1GU
Standards
ISO-2813,ISO-7688,GB8807
Operating language
Chinese
Dimensions
114*35*65mm
Weight
300g
Power
1 AA battery
Stability
±2GU
Measurement port size
14*28mm
Working Voltage
1.5V
Application
The main test metal can also measure Coatings, Paints, Printing inks, Plastics, Stone, Flooring, Woodware in the range of 0-199
Measurement port shape
ellipse
JFL-B60M (common type for metal)
Glossmeter
Metal Surface Glossiness Detection Instrument JFL-B60M JFL 60 Degree Glossiness Detection Instrument Picture

Introduction

Jinfulan JFL-B60M Metal Surface Gloss Tester is a testing instrument with a projection angle of 60 degrees and is used to measure the gloss of metal surfaces. The measuring range is 0~199.9Gs. JFL-B60M Metal Surface Gloss Tester is a high-precision, miniaturized gloss Tester manufactured in accordance with the ISO-2813 standard. It is suitable for specular gloss testing of coatings, paints, inks, plastics, stones, floors, and wooden furniture surfaces. The measurement conforms to ISO, ASTM and other domestic (CN) and foreign standards.

Applications

JFL-B60M Metal Surface Gloss Tester is mainly used to measure metal surface gloss, and can also be used to measure paint, paint, ink, plastic, stone, floor, wood, etc. in the range of 0~199Gs.

product structure

JFL JFL-B60M metal surface gloss Detector structure diagram

operation method

1. Press the switch of the JFL-B60M Metal Surface Gloss Tester , place the JFL-B60M Metal Surface Gloss Tester on the black glass calibration plate, and turn the adjustment knob to make the displayed value the same as the nominal value of the standard plate.

2. Put the JFL-B60M metal surface gloss Detector on the white ceramic standard plate, and the error between the displayed value and the nominal value of the white ceramic standard plate should not be greater than ±1.2 gloss units, otherwise, the standard plate is dirty and other issues should be considered .

3. When the error is out of tolerance, calibrate again according to the previous two steps.

4. Measure the sample after calibration, when the measured value exceeds 199.9 gloss units, the display overflows "1".

Pictures

JFL JFL-B60M metal surface gloss Detector real shot

FAQ
QWhy JFL-B60M surface gloss detector in the white calibration plate measurement error is 1.5GS?
A Hello, 1.5Gs is beyond a reasonable error range, please check if the standard board is dirty.
QI want to do a test on the ASTMF1843 standard, is this glossmeter okay?
A Sure
QWhy has my JFL-B60M gloss detector been used for many years, and now when calibrating, the error of the nominal value is always greater than ± 1.2 Gloss units?
A Hello, according to what you said, there are several possibilities. First check if the standard version is dirty. The problem.