China

Glossiness Meter JFL-BZ60M JFL Detection of Metal Surface Gloss Instrument

SE
The JFL-BZ60M Glossiness Meter is a smart Glossiness Meter with a projection angle of 60 °. The JFL-BZ60M Glossiness Meter is mainly used for the measurement of Glossiness on metal surfaces, and can also be used for paints, coatings, plastics and other Gloss tests.
$316.00/set
PromotionLoading...
Coupon
Certificate

JFL JFL-BZ60M (metal smart) GlossmeterSPEC

JFL JFL-BZ60M (metal smart) GlossmeterDetails

JFL JFL-BZ60M (metal smart) GlossmeterPacking list

SKU
NB001299
Geometry
60°
Measurement range
0~3000GU
Measuring Areas
10*20mm
Error of indication
±2GU
Division Value
1GU
Standards
ISO-2813、ISO-7668、GB8807
Operating language
Chinese
Dimensions
114*35*65mm
Weight
300g
Power
Rechargeable battery
Stability
±2GU
Measurement port size
14*28mm
Working Voltage
4.8V
Application
High Gloss Display Range) The main test metal can also measure Coating, Paint, Printing ink, Plastic, Stone, Floor, Wood in the range of 0-199
Measurement port shape
ellipse
JFL-BZ60M (metal smart)
Glossmeter
Glossiness Meter JFL-BZ60M JFL Detection of Metal Surface Gloss Instrument Picture

Introduction

Jinfulan JFL-BZ60M gloss meter is an intelligent gloss meter for metals, used to measure the gloss of metal surfaces, using a projection angle of 60°, the display range is 0~3000Gs, and the measurement error is <2Gs.

JFL-BZ60M gloss meter can basically meet the measurement of all metal high-gloss surfaces, and is also suitable for the measurement of surface gloss units of coatings, paints, plastics and other products in the range of 0-199 ordinary gloss. The gloss meter has the characteristics of small size, easy to carry, accurate and stable measurement.

structure diagram

Jinfulan JFL-BZ60M gloss meter structure diagram


Calibration steps

1. Put the measurement window of the JFL-BZ60M gloss meter on the black velvet board, press the "set number/zero" key, and the displayed value is 0.0.

2. Use the black glass standard board instead of the black velvet board, put it under the measurement window of the JFL-BZ60M gloss meter , press the "calibration/measurement" button, enter the setting and loading, and then display "000.0" and the first "0" flashing.

3. Press the "Shift" key to move the flickering digit to the right, and press the "Number/Zero" button to change the value of the "blinking" digit from 0-9-0. By pressing the "Shift" and "Number/Set" "Zero" key causes the displayed value to be the same as

The nominal value of the black glass standard plate is the same, which means that the setting by pressing the "Calibration/Measurement" button is completed .

4. Put the JFL-BZ60M gloss meter measurement window on the white ceramic standard plate, the error should not be greater than ±1.2 gloss units, otherwise the standard plate is dirty and other problems should be considered .

5. After the JFL-BZ60M gloss meter is calibrated, the gloss of the sample can be measured.

Precautions

When the power is insufficient, the word "bAtt" will be displayed. At this time, it should be charged . The charging time is about four hours. Charging standard "⊕".

Pictures

Jinfulan JFL-BZ60M Gloss Meter Real Shot


FAQ
QWhat is the reason why my gloss meter shows bAtt?
A Hello, this is the display words of Jinfulun Gloss meter insufficient power, it needs to be charged in time.
QWhat material is the shell of JFL-B60 Gloss meter?
A Hello, this is an all-metal body.
QCan this JFL-BZ60M be used for the Gloss test of metallic paint?
A Hello, the gloss meter can basically test the Gloss of the paint, but it depends on what the Gloss range of your paint is, is it high-gloss or matte?
QWhat size is the measurement port of Jinfulun Gloss meter small hole model?
A Hello, JFL-B60MS is a small hole model, and the measurement window size is 1.5x3mm.
QCan the JFL JFL-BZ60M (intelligent for metal) glossmeter meet the JJG696 standard?
A Applicable, you can consult the customer service of the mall for details about the JJG696 standard.
Standard