• 8- and 12-channel pipettes of various volumes for standard 96-well plates;
• The nozzle ejector can push out multiple nozzles at the same time, which is effective and labor-saving;
• The lower half of the pipette can be rotated 360 degrees for easy pipetting;
• Each nozzle has an independent piston device, which is convenient for maintenance and maintenance;
• Special nozzle connection design, easy to observe the sealing condition of the nozzle;
• Online calibration available.

| Range | increment | Measure the volume | Maximum systematic error is allowed (Inaccuracy) | The maximum random error is allowed (Inaccuracy) | ||
| unit | μl | % | μl | % | μl | |
| 0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
| 5 | 2.50% | 0.125 | 2.50% | 0.125 | ||
| 1 | 4.00% | 0.04 | 4.00% | 0.04 | ||
| 5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
| 25 | 1.50% | 0.375 | 1.00% | 0.25 | ||
| 5 | 3.00% | 0.15 | 2.00% | 0.1 | ||
| 50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
| 150 | 1.00% | 1.5 | 0.50% | 0.75 | ||
| 50 | 1.50% | 0.75 | 0.80% | 0.4 | ||