SDPTOP DCY-1015 Laboratory Low Temperature Thermostat -10~95 ℃/15L
SE
Temperature range -10~95 ℃, Temperature fluctuation +/- 0.05 ℃, temperature transducer Pt100, temperature resolution 0.1 ℃, working tank volume 15L, pump flow 10L/min, newly designed microcomputer precision thermostatic control system and imported Pt100 temperature transducer
Temperature range -10~95 ℃, Temperature fluctuation +/- 0.05 ℃, temperature transducer Pt100, temperature resolution 0.1 ℃, working tank volume 15L, pump flow 10L/min, newly designed microcomputer precision thermostatic control system and imported Pt100 temperature transducer
Features
New design microcomputer precision constant temperature control system Imported Pt100 temperature sensor
Well-designed and thoughtful refrigeration system, high refrigeration efficiency, fast cooling, low noise and small vibration
Precise and reasonable design of dual-compressor cascade refrigeration system and heat exchange system ensures reliable operation and rapid cooling of the instrument
Humanized design: Foldable handles on both sides make it easy to move the instrument. DC30 series and constant temperature baths below -40°C use heavy-duty self-locking casters for easy movement.
The low-temperature tank adopts a professional overall foaming process, which has good heat insulation performance and effectively reduces the loss of cooling capacity
The professional circulating stirring design ensures good horizontal and vertical uniformity of the liquid in the tank. The volume of the low-temperature tank above 20L is 10L/min, and the remaining 8L/min.
The inner tank and box are made of high-quality stainless steel (DC0504B)
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
QHow to choose the liquid medium in the Thermostatic bath?
A
When the Operating temperature is 6~ 85 ℃, the liquid medium is pure water or distillation water; when 85~ 95 ℃, 15% glycerol aqueous solution can be used; when the Operating temperature is lower than 6 ℃, the liquid medium is generally ethanol, ethylene glycol, silicone oil (viscosity ≤ 50mPa · s) or Miscellaneous freezing point is lower than the Operating temperature non-corrosive medium.
QMust the liquid in the tank be higher than the evaporation tube?
A
The liquid medium is added to the tank to immerse the evaporation tube in the medium, and the liquid level is about 20 to 30mm lower than the working platform. During use, the medium will slowly evaporate and the liquid level will drop. When the liquid level is lower than the evaporation tube, the medium should be added in time. In particular, it should be noted that the external circulation of the medium in the tank into the thermostatic container will reduce the liquid level in the tank and expose the evaporation tube to the liquid level. It is forbidden to power on the liquid medium in the tank.
QWhat is the limit low temperature of this DCY-1015 Low Temperature Thermostatic bath?
A
SDPTOP DCY-1015 meets the Temperature range of -10 to 95 ℃, with a minimum temperature of -10 ℃.
QHelp me find a low temperature Thermostat that meets the JB/T5376 standard
A
This DCY-1015 Low Temperature Thermostat is fine!