China

MODERNER NDJ-8S Digital Rotational Viscometer viscosity range is 1-2000000cp

SE
Digital display, measurement range 1~ 2000000mPa · s, rotor specifications 1 #, 2 #, 3 #, 4 #, Rotor speed 0.3/0 6/1.5/3/6/12/30/60rpm, reproducibility +/- 0.5% F.S, measurement accuracy +/- 2% F.S.
Model:
Certificate

Moderner NDJ-8S Digital Rotational ViscometerSPEC

Moderner NDJ-8S Digital Rotational ViscometerDetails

Moderner NDJ-8S Digital Rotational ViscometerPacking list

SKU
NB001455
Display
digital display
Measuring range
1~2000000mPa·s
Rotor Specifications
1#、2#、3#、4#
Rotor speed
0.3/0.6/1.5/3/6/12/30/60rpm
Reproducibility
±0.5%F.S
Measurement Accuracy
±2%
Power
110~230V 50Hz/60Hz
Interface
One R232C communication interface, one temperature probe interface, and one power jack.
Overall Dimensions
300*300*450mm
Weight
5kg
Display information
Automatically displays the maximum Viscosity value for measurement of the selected rotor and rotational speed combination
Automatic scan function
Viscosity of the test sample can automatically prompt the best combination of Rotor velocities
NDJ-8S
Digital Rotational Viscometer
Digital display, measurement range 1~ 2000000mPa · s, rotor specifications 1 #, 2 #, 3 #, 4 #, Rotor speed 0.3/0 6/1.5/3/6/12/30/60rpm, reproducibility +/- 0.5% F.S, measurement accuracy +/- 2% F.S.
MODERNER NDJ-8S Digital Rotational Viscometer viscosity range is 1-2000000cp Picture

Introduction to NDJ-8S Digital Display Rotational Viscometer

NDJ-8S digital display Rotational Viscometer is a digital display Rotational Viscometer designed according to the JJG-2005 standard in MODERNER. Compared with NDJ-5S Rotational Viscometer, it has a larger viscosity range. Equipped with No. 1, No. 2, No. 3 and No. 4 rotors. When the viscosity of the measured sample is more than 15 cp, it is necessary to purchase No. 0 low-viscosity rotor for use. NDJ-8S digital display rotary viscometer adopts the latest international microcomputer control technology, excellent imported stepping motor, its LCD with backlight function, through program setting, automatic speed control, with good accuracy, stable speed , Easy to operate, automatic scanning, error alarm and other practical functions.

Key Function

1) ▲/▼ function keys, used to change the rotation speed and select the rotor number

2) "Rotor Selection" function key, prompting the rotor number, the cursor flashes

3) "Speed speed setting" function key, prompting the speed, the cursor flashes

4) "Start, Stop" function key to execute or terminate the current measurement program

5) "Auto scan" function key, for unknown samples to help users choose the appropriate rotor speed measurement

6) "OK" function key to confirm the currently selected information

Features

1) NDJ-8S Digital Rotational Viscometer uses standard silicone oil to measure and calibrate the full scale and linearity of the viscometer through the PC interface. It has the characteristics of high sensitivity, high precision and high repeatability.

2) LCD display content: rotor number, speed, range, viscosity, temperature, opening angle %.

3) With the function of % of the measured value in the full scale, it is very convenient for the user to select the rotor and rotational speed for accurate measurement of the sample liquid.

4) The stepping machine intelligently adjusts the speed according to the program, abolishing the old method of gear speed adjustment.

5) It has the advantages of stable speed, quick and easy operation, high measurement accuracy, good anti-interference performance, and wide working Voltage (110V, 220V, 50Hz, 60Hz are all available).

Application

NDJ-8S Digital Rotational Viscometer is suitable for measuring solvent-based adhesives, latex, biochemical products, paints, coatings, cosmetics, inks, pulp, food, starch, etc.

Principle of Rotational Viscometer

MODERNER NDJ-8S digital display Rotational Viscometer Principle diagram

As shown in the figure above, the stepping motor is driven with high resolution to drive the sensor pointer, and the rotor is driven to rotate through the hairspring and the rotating shaft. If the rotor is not resisted by the fluid, the hairspring sensor pointer is in the same position as the stepper motor sensor pointer. Conversely, if the rotor is subjected to the viscous resistance of the liquid, the balance spring will generate torque to counteract the viscous resistance, and finally reach a balance. At this time, the output signal of the photoelectric sensor is sent to the 16-bit microcomputer processor for data processing, and finally the viscosity value (mPa·s) of the liquid is displayed on the LCD with night vision function.

FAQ
QI want to measure the viscosity of paints, which Viscometer is suitable?
A The NDJ-8S digital rotational viscometer is recommended. This instrument has a wide range of viscosity measurement, from 1 to 2 million mPa · s, and fully covers the viscosity range of paints. It is equipped with an automatic scanning function that also helps you quickly find the most suitable rotor and speed combination. Simple operation and high measurement accuracy are suitable for viscosity detection of samples such as paints.
QWhat configuration should I choose for low viscosity samples (such as below 15cp)?
A If you want to measure low viscosity samples below 15cp, it is recommended to choose the NDJ-8S with an additional 0 low viscosity rotor. The standard 1-4 rotor is suitable for higher viscosity, while the 0 rotor is specially designed for low viscosity samples to ensure measurement accuracy and stability in the low viscosity range to meet your testing needs for low viscosity samples.
QCan this Viscometer test the viscosity of real stone paint? Within 20W
A OK
QEpoxy resin Coating viscosity measurement with what material Viscometer is better?
A Measurement Coating viscosity generally can choose viscosity cup or Viscometer, viscosity cup mainly measures relative viscosity, Viscometer measures dynamic viscosity. In general, in addition to thixotropic Coating, Coatings under 150s can be coated with 4 cups, and Coatings under 20 can be coated with 1 cup. In addition, Yehui uses ISO flow cup to measure Newton-type liquid Coating viscosity, Stormer Viscometer to measure architectural coatings viscosity, and for all brush coatings, rotational viscometer or Cone and Plate Viscometer measurement can be used.
QMeet GB/T12004.4 digital rotational viscometer
A NDJ-8S digital rotational viscometer, please refer to it.
QCan a 5 #rotor be added to the Viscometer of this model NDJ-8S? How much is the price?
A The No. 4 rotor is already the smallest, and there is no No. 5 rotor. If you need to add additional rotors, only No. 0 is available.
QWhat is the difference between the NDJ-8S rotational viscometer and the NDJ-5S?
A Hello, both are rotational viscometers. The NDJ-8S rotational viscometer will be larger in the viscosity measurement range, which is 1-2000000cp.
Standard