1: Specially used to measure very low viscosity samples
2: The sample volume only needs 25ml
3: The cylindrical rotor can accurately calculate the shear rate, which is convenient for detailed analysis of products.
1. For the liquid below 15mpa.s, choose the No. 0 rotor, use the No. 0 rotor without the protective frame, and install the No. 0 rotor on the connecting screw (rotate counterclockwise to install).
2. Insert the fixing sleeve into the lower cylinder of the instrument, and tighten it with the sleeve fixing screw.
3. Pour 20-25ml of the liquid to be tested into the outer test tube (with bottom).
4. Put the outer test tube into the fixing sleeve and tighten it with the test tube fixing screw. When tightening, pay attention to the tapered end of the test tube fixing screw so that it can be screwed into the triangular groove on the upper end of the outer test tube (it can be inserted into the round hole on the side) Observe whether the triangular groove of the test tube is in the center of the round hole). The test can be carried out after controlling the temperature of the liquid to be tested.