| PH range | -2.000~20.000pH |
| PH measurement Accuracy | ±0.002pH |
| resolution | 0.01/0.001pH/optional |
| Temperature Measurement range | 0~100℃/32~212℉ |
| Power | DC12V power adapter |
| Dimensions | 240*200*80mm |
| Weight | 1.7kg |
| PH calibrated | 1 to 5 point |
| Calibrating fluid | USA, NIST, DIN, custom |
| mV range | -2000~2000mV |
| mV measurement Accuracy | ±0.2mV |
| mV resolution | 0.1mV |
| ORP calibrated point | 1 point (only for R.mV mode) |
| ORP measurement mode | Relative or absolute mV |
| Temperature Measurement range | 0~105℃/32~221℉ |
| Temperature measurement Accuracy | ±0.5℃ |
| temperature resolution | 0.1℃ |
| Temperature calibrating point | 1 point |
| Temperature unit | ℃, ° F, optional |
| data lock | Manual or automatic end point locking |
| Data stability condition | Standard or high precision |
| Calibrated reminder | 1 to 31 days or closed |
| Interval reading | 10, 30, 60 seconds, 10, 30 minutes or off |
| One-click reset function | contain |
| data storage | 1000 groups |
| Communication interface | USB |
| Connector | BNC |
| Display | 7-Inch TFT LCD |
| Optional accessories | P11 type pH electrode, P11-LiCl type pH electrode, P11-HA type pH electrode, P11-NA type pH electrode |
| Optional accessories | P12 type pH electrode, P13 type pH electrode, P15 type pH electrode, P16 type pH electrode, P18 type pH electrode, P19 type pH electrode, P21 type pH electrode, P22 type pH electrode |
| Optional accessories | TP-10K temperature probe, 501 type ORP electrode, 502 type ORP electrode, pH standard buffer reagent (USA standard), pH standard buffer reagent (NIST standard) |
| Packaging Dimensions | 300*240*170mm |
| Encasement Weight | 2kg |
| Ion Measurement range | 0~30000Ppm, mg/L, mol/L (please refer to the ion electrode description for the actual range) |
| Ion measurement Accuracy | ± 0.5% F. S (primary ion), ± 1% F. S (secondary ion) |
| Ion display resolution | 0.001, 0.01, 0.1, 1 |
| Ion calibrated point | 2 to 5 point |
| Ion measurement mode | Direct reading method, known addition method, known reduction method, sample addition method, sample reduction method |
| Hardness Measurement range | 0.05~200mmol/L |
| Hardness measurement Accuracy | ±1% F.S |
| Water Hardness Display Resolution | 0.001, 0.01, 0.1, 1 |
| Hardness calibrated point | 2 to 5 point |
| Water Hardness Measurement Units | ° dH, ° e, ° fH, mg / L, mmol / L |
| Temperature compensation method | Automatic/Manual |