Bante900-CN laboratory multiparameter Water Quality Meter

SE
High-precision multiparameter Water Quality Analyzer, including 8 measurement functions, 15 function parameters can be customized in the setting menu, the meter is suitable for laboratory measurement of pH, ORP, Ion concentration, conductivity, TDS, salinity, resistivity, dissolved oxygen concentration and saturation
$778.00/set
PromotionLoading...
Coupon
Model:
Certificate

Bante Bante900-CN Laboratory multiparameter Water Quality MeterSPEC

Bante Bante900-CN Laboratory multiparameter Water Quality MeterDetails

Bante Bante900-CN Laboratory multiparameter Water Quality MeterPacking list

SKU
NB014822
PH range
-2.000~20.000pH
Measurement Accuracy of pH
±0.002pH
pH resolution
0.1, 0.01, 0.001 pH, optional
pH calibrating point
1 to 5 point
pH calibrating solution
USA, NIST, DIN, Custom
pH temperature compensation range
0~100℃, 32~212°F
pH temperature compensation
Manual/Automatic
ORP Measurement range
±1999.9mV
Measurement Accuracy of mV
±0.2mV
mV resolution
0.1, 1mV, optional
mV calibrated point
1 point (only for R.mV mode)
mV calibrated range
±200mV
Measurement range
0.001~ 19999ppm, mg/L, mol/L (please refer to the ionic electrode description for the actual range)
Measurement Accuracy
+/- 0.5% F. S (diion ), +/- 1% F. S (diion)
Ion resolution
0.001, 0.01, 0.1, 1
Ion calibrated point
2 to 5 point
Ion calibrating fluid
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm, mol/L, mg/L
Ion temperature compensation range
0~100℃, 32~212°F
Ion temperature compensation
Manual or automatic
Conductivity range
0~20mS/cm (最大 200mS/cm)
Measurement Accuracy of conductivity
±0.5%FS
EC resolution
0.001, 0.01, 0.1, 1
Conductivity calibrated point
1 to 5 point
Conductivity calibrating fluid
10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
EC Temperature Compensation Mode
Manual or automatic
EC Temperature Compensation Coefficient
0.0~10.0%/℃
EC Temperature Compensation Mode mode
Linear or pure water
Conductivity Detector pool constant
K = 0.1, 1, 10, custom
Standardized temperature
20 ℃, 25 ℃, optional
EC Temperature Compensation Range
0~100℃, 32~212°F
TDS range
0~10ppt (maximum 200ppt)
TDS Measurement Accuracy
±1%FS
TDS conversion factor
0.1~ 1.0 (default 0.5)
Salinity range
0~10ppt (maximum 80ppt)
Measurement Accuracy of salinity
±1%FS
Measurement mode of salinity
Utility salinity (psu) or seawater salinity (ppt)
Resistivity range
0 ~ 10MΩ (? 20MΩ)
Resistivity range
±1% F.S
Resistivity resolution
0.01, 0.1, 1
DO range
0.00~20.00mg/L
DO Measurement Accuracy
±0.2mg/L
DOS resolution
0.1, 0.01 mg/L, optional
DO temperature compensation range
0~50℃
DO temperature compensation
automatic
Pressure compensation range
60.0~112.5kPa, 450~850mmHg
Salinity compensation range
0~50g/L
DO% Measurement range
0.0~200.0%
DO% Measurement Accuracy
±2.0%
DO% resolution
0.1%
Measurement range
0~105℃, 32~221℉
Measurement Accuracy
±0.5℃
Temperature Resolution
0.1℃
Temperature calibrating point
1 point
Temperature unit
℃, ° F, optional
Data stability condition
Fast or high-precision
Calibrated reminder
1 to 31 days or closed
shutdown mode
Manual or automatic (10, 20, 30 minutes after no button operation)
Data storage
500 sets
Communication interface
USB
Connector
BNC, 6-pin MINI Connector
Display
Customized LCD screen (130 * 110mm)
Power Supply
DC5V power adapter
Dimensions
210*188*60mm
Instrument Weight
1.5kg
Optional Accessories
P11 type pH electrode, P11-LiCl type pH electrode, P11-HA type pH electrode, P11-NA type pH electrode
Optional Accessories
P12 type pH electrode, P13 type pH electrode, P15 type pH electrode, P16 type P22 type pH electrode
Optional Accessories
501 type ORP electrode, 502 type ORP electrode, DO100 type dissolved Oxygen electrode, CON-10 type Conductivity Detector electrode, CON-0.1 type Conductivity Detector electrode, CON-1 type Conductivity Detector electrode,
Optional Accessories
NO3-US nitrate ion electrode, Ca-US calcium ion selective electrode, F-US fluoride ion selective electrode, Na-US sodium ion selective electrode, CL-US chloride ion selective electrode, CL-US chloride ion selective electrode,
Optional Accessories
Pb-US lead ion selective electrode, I-US iodine ion selective electrode, Cu-US copper ion selective electrode, Cd-US cadmium ion selective electrode, Br-US bromine ion selective electrode, NH4-US ammonium ion selective electrode, Cn-US cyanide ion selective electrode,
Optional Accessories
NH3-US Ammonia Gas sensing electrode, S-US sulfur ion selective electrode, Ag-US silver ion selective electrode, K-US potassium ion selective electrode, S-US sulfur ion selective electrode,
Optional Accessories
TP-10K temperature probe, dissolved Oxygen electrode membrane cap, dissolved Oxygen electrode filling solution, Conductivity Detector standard solution, ion standard solution, ion strength adjuster, pH standard buffer reagent (USA standard), pH standard buffer reagent (NIST standard)
Packaging Dimensions
300*240*170mm
Encasement Weight
1.7kg
Bante900-CN
Laboratory multiparameter Water Quality Meter
High-precision multiparameter Water Quality Analyzer, including 8 measurement functions, 15 function parameters can be customized in the setting menu, the meter is suitable for laboratory measurement of pH, ORP, Ion concentration, conductivity, TDS, salinity, resistivity, dissolved oxygen concentration and saturation
Bante900-CN laboratory multiparameter Water Quality Meter Picture

Bante900 multi-parameter Water Quality Analyzer

Multi-parameter laboratory TDS meter with built-in system menu, 15 functional parameters to choose from, the instrument is suitable for measuring pH, redox potential, ion concentration, conductivity, total dissolved solids (TDS), salinity, resistivity, dissolved oxygen concentration and saturation of liquids.

pH Mode Features

• Multi-parameter Water Quality Analyzer with large backlit LCD screen.

• 1 to 5 point push-button calibration for automatic recognition of pH standard buffers.

• Optional pH buffer standard (USA/NIST/DIN) or 2-point custom calibration.

• Automatic electrode slope display to prompt the user whether to change the pH electrode.

ORP mode features:

• 1-point offset calibration that allows adjustment of the display value to a known standard.

• Relative and absolute mV modes provide accurate ORP readings.

Ion Mode Features

• 2 to 5 pushbutton calibration, including 8 concentration calibration points.

• Automatic identification of ion-selective electrodes without the need to specify the ion type.

• Direct reading of ion concentration values to avoid tedious conversions and calculations.

• Concentration units (ppm, mg/L, mol/L) are available to meet a variety of ion measurement needs.

TDS meter

TDS meter

Conductivity/TDS/Salinity/Resistivity Mode Features

• 1 to 5 point key calibration for automatic identification of conductivity standards.

• Selectable cell constant (K=0.1/1/10), TDS conversion factor, normalized temperature (20 or 25°C), linear and pure water temperature compensation,

  Seawater and practical salinity measurement mode.

Dissolved Oxygen Mode Features

• 1 or 2 button calibration, using air saturated water or zero oxygen solution.

• Salinity and atmospheric pressure compensation to improve the measurement accuracy of the instrument.

General features:

• Automatic temperature compensation for accurate measurements over the full scale.

• Automatic calibration reminder to remind the user to calibrate the meter regularly.

• Automatic endpoint locking, stable measured values are retained for easy browsing and recording.

• Automatic electrode diagnostics with zero offset, electrode slope, range and conductivity K values.

• The system menu can customize 15 functional parameters such as the number of calibration points, stability standard, data lock mode, automatic shutdown, etc.

• One-key Reset automatically restores the meter to factory default settings.

• Extended memory for up to 500 sets of measurement data.

• Built-in clock automatically marks the date and time of stored data to comply with GLP specifications.

• USB communication interface to send stored data to a personal computer.

FAQ
QWhat does the screen display PURE mean?
A It refers to the pure water compensation mode.
QConductivity detector electrode how to clean stains?
A Immerse the electrode Sensors in tap water for 10 minutes to remove oil or stains that may be attached to the Platinum Sensors.
QHow to calibrate STIP-scan temperature?
A Specific operations: 1. Ensure that the temperature probe is connected to the meter; 2. Press and hold the ℃ key for 3 seconds, the meter enters the temperature calibrate mode, and the screen displays the current temperature value; 3. Press up or down to set the temperature value; 4. Press the enter key to confirm, the calibrate is complete.
QThe screen displays --- what is the reason?
A There are two possible reasons: 1. The electrode Sensors are too drying, which can be solved by immersing in tap water for 10 minutes; 2. Measured value Excessive range: Check whether the Sensors are clean and not contaminated.
Standard