MODERNER CNDB-3 Zahn Viscosity Cup (zahn cup) 3.8mm Zahn cup

SE
Viscosity range 150~850cSt, outflow time 20~80s, flow hole diameter 3.8mm, capacity 45ml, cup body material stainless steel, filter material stainless steel, handle material stainless steel, applicable standard ASTM D4212-93.
Model:
Certificate

Moderner CNDB-3 Zahn Viscosity CupSPEC

Moderner CNDB-3 Zahn Viscosity CupDetails

Moderner CNDB-3 Zahn Viscosity CupPacking list

SKU
NB015114
Viscosity range
150~850cSt
outflow time
20~80s
Flow hole diameter
3.8mm
Capacity
45ml
Cup body material
stainless steel
Filter material
stainless steel
Handle material
stainless steel
Applicable standard
ASTM D4212-93
CNDB-3
Zahn Viscosity Cup
Viscosity range 150~850cSt, outflow time 20~80s, flow hole diameter 3.8mm, capacity 45ml, cup body material stainless steel, filter material stainless steel, handle material stainless steel, applicable standard ASTM D4212-93.
MODERNER CNDB-3 Zahn Viscosity Cup (zahn cup) 3.8mm Zahn cup Picture

use

Zahn Viscosity Cups are used to test the viscosity of Newtonian or near-Newtonian fluids. CNDB type Chai's Viscosity Cup is manufactured according to ASTM D4212-93 standard.

Model Selection

model

Measuring range

Recommended outflow time

Aperture

CNDB-1

5-6035-802.0mm

CNDB-2

20-23020-802.7mm

CNDB-3

150-85020-803.8mm

CNDB-4

220-110020-804.3mm

CNDB-5

460-184020-805.3mm
Unit: cSt, s, mm
Other parameters
volume

45± 2ml


zahn portable Viscosity Cup

Use and maintenance

1. Clean the cup body with a suitable solvent before and after use.

2. Choose an appropriate cup number, and try to make the outflow time meet the recommended outflow time in the above table.

3. Immerse the cup body in the container for 1 to 5 minutes to make the cup reach thermal equilibrium, then pull the cup vertically out of the raw material, and the action needs to be fast and steady. When the top edge of the cup leaves the liquid surface, keep the cup body vertical when the liquid flows, and the distance between the bottom of the cup and the liquid surface of the container should not exceed 150mm. Stop timing when the first breakpoint occurs, and the number of seconds of outflow time at this time indicates a certain viscosity value. The difference between the two measured values should be less than 5% of the average value.

4. Convert the outflow time into kinematic viscosity according to the following formula: (sample temperature is 25°C±0.5°C when measuring)

    ν = k(t - c)

    In the formula: ν—kinematic viscosity, cSt

           t — outflow time, s

           k, c — corresponding constants (see table below)

cup number

1

2

3

4

5

k

1.1

3.5

11.7

14.8

twenty three

c

29

14

7.5

5

0

FAQ
QThe main use of CNDB Zahn viscosity cup?
A CNDB Zahn viscosity cup is mainly used to test the viscosity of Newtonian or approximate Newtonian fluids.
QWhat steps should I pay attention to when using the CNDB Zahn viscosity cup?
A The cup needs to be cleaned with a suitable solvent before and after use. Select the appropriate cup number to make the outflow time meet the recommended time as much as possible. After immersing the cup body in the container for 1 to 5 minutes to achieve hot equilibrate, then pull out the cup vertically. The action should be fast and smooth. From the time the upper edge of the cup leaves the liquid surface, the liquid should be kept vertical when flowing, and ensure that the distance between the bottom of the cup and the liquid surface of the container does not exceed 150mm. Stop timing when the first breakpoint occurs. At this time, the number of seconds of outflow time indicates a certain size of viscosity value. The difference between the two measured values should be less than 5% of the average value.
QCan ZAHN viscosity cup be used to test the viscosity of emulsion paint?
A Hello, it's OK.
QDoes this Zahn viscosity cup meet the JJG743 standard?
A The Zahn viscosity cup is suitable for JJG743, please consult customer service for detailed application definition
QEpoxy resin Coating viscosity measurement with what material Viscometer is better?
A Measurement Coating viscosity generally can choose viscosity cup or Viscometer, viscosity cup mainly measures relative viscosity, Viscometer measures dynamic viscosity. In general, in addition to thixotropic Coating, Coatings under 150s can be coated with 4 cups, and Coatings under 20 can be coated with 1 cup. In addition, Yehui uses ISO flow cup to measure Newton-type liquid Coating viscosity, Stormer Viscometer to measure architectural coatings viscosity, and for all brush coatings, rotational viscometer or Cone and Plate Viscometer measurement can be used.
QPrinting ink printing viscosity measurement, this one applies?
A Yes, Printing ink printing general viscosity can be applied to No. 3 zahn cup
Standard