3NH YS6010 Benchtop Colorimeter 400nm Cut-off Light source

SE
TFT true color 7inch capacitive touch screen, Full Light source, Lighting mode: reflection D/8 °, transmission D/0 ° (including UV/exclude UV measurement) Color measurement accurate and stable, Storage capacity, PC side powerful expansion function, for laboratory Color accurate analysis and transmission.
Model:
Certificate

3NH YS6010 Desktop ColorimeterSPEC

3NH YS6010 Desktop ColorimeterDetails

3NH YS6010 Desktop ColorimeterPacking list

SKU
NB015542
Lighting method
Reflection: D/8 (diffuse illumination, 8 ° direction reception); SCI/SCE measurement; including UV/exclude UV measurement; Transmission: D/0 (diffuse illumination, 0 ° direction reception) SCI/SCE measurement; including UV/exclude UV measurement
Sensors
256-Pixel dual array CMOS image sensor
lighting source
360~780Nm combined LED light source, 400nm cut-off light source
Measurement time
About 2.4s (about 5s for simultaneous SCI/SCE testing)
Measurement caliber
反射:Φ30mm/Φ25.4mm,Φ10mm/Φ8mm,Φ6mm/Φ4mm;透射:Φ30mm/Φ25.4mm
Operating language
Simplified Chinese, Traditional Chinese, English
Data Interface
USB, print serial port
Viewing Angle
CIE 2 ° / 10 °
Observation light source
D65,A,C,D50,D55,D75,F1,F2,F3,F4,F5,F6,F7,F8,F9,F10,F11,F12
Color difference formula
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99,ΔE(Hunter)
Data storage
2000 master standards and 20,000 samples (SCI/SCE counts as one data)
Operating conditions
0~40℃(32~104°F)
Storage environment
-20~50℃
Dimensions
370*300*200mm
Weight
Approx. 9.6kg
Standards
IE No.15,GB/T 3978,GB 2893,GB/T 18833,ISO7724/1,ASTM E1164,DIN5033 Teil7
Integrating sphere size
F154mm
Spectroscopic method
Concave grating spectroscopy
Wavelength range
360~780nm
wavelength spacing
10nm
Light spectrum half bandwidth
10nm
Reflectance Measurement Range
0~200%
Light-containing method
Reflection SCI/SCE, Transmission SCI/SCE
Color space
CIE LAB,XYZ,Yxy,LCh,CIE LUV,Musell,s-RGB,HunterLab,βxy,DIN Lab99
Chromaticity indicator
WI, YI, TI, homochromatic index MI, dip lightfastmess, color fastness, force, coverage, cobalt-platinum index, Gardner index, 8 degree glossiness, 555 hue classification
display mode
Light spectrum graphs/data, sample chromatic values, color difference values/graphs, chromaticity graphs, Color simulation, Pass/Fail results, Color bias
Measurement time
About 3s (about 6s for simultaneous SCI/SCE testing)
Spectral reflectance repeability
Φ25.4mm/SCI, standard deviation within 0.05% (400~700nm: within 0.04%)
Chromatic value repeability
Φ25.4mm / SCI, ΔE * from 0.02<--sep--0/> <--sep--0/> Φ25.4mm / SCI, ΔE * from 0.03<--sep--0/> <--sep--0/>
Spectral transmittance, sexual repeability
Φ25.4mm/SCI, standard deviation within 0.05% (400~700nm: within 0.04%)
Inter-station difference
ΔE * from < 0.15
Power
DC 24V, 2A power adapter Power
Lighting source life
5 years more than 3 million measurements
YS6010
Desktop Colorimeter
TFT true color 7inch capacitive touch screen, Full Light source, Lighting mode: reflection D/8 °, transmission D/0 ° (including UV/exclude UV measurement) Color measurement accurate and stable, Storage capacity, PC side powerful expansion function, for laboratory Color accurate analysis and transmission.
3NH YS6010 Benchtop Colorimeter 400nm Cut-off Light source Picture

YS60 series is an independently developed domestic (CN) desktop grating Spectrophotometer with independent intellectual property rights, TFT true color 7inch capacitive touch screen, full light source, lighting method: reflection D/8°, transmission D/0° (including UV/excluding UV Measurement) Accurate and stable color measurement, large storage capacity, and powerful expansion functions on the PC side, used for accurate analysis and transmission of color in the laboratory.

Features

1. High hardware configuration: 7-inch TFT pure color capacitive touch screen; Bluetooth 2.1; concave grating;

2. Dual-array 256-pixel CMOS Detector; long-life and stable LED/ultraviolet LED/xenon lamp;

3. Measure the reflection and transmission spectrum of the sample, and the Lab data is accurate, which can be used for color matching and accurate color transfer;

4. Caliber automatic identification, four calibers of φ25.4/15/8/4mm can be switched arbitrarily, while taking into account the special needs of customers;

5. Temperature monitoring and compensation, built-in temperature sensor, monitors and compensates the test environment to ensure more accurate measurement results;

6. The test Wavelength range is 360~780nm, and built-in 400nm cut-off/420nm cut-off/460nm (limited to xenon lamp version), the test UV is more professional;

7. The independent light source Detector monitors the change of the light source at all times to ensure the reliability of the light source;

8. A variety of measurement modes: quality control mode, sample mode, to meet the needs of more customers;

9. A variety of accessories: reflective sample holders, transmission mounts, and more Operating Conditions;

10. Large-capacity storage space, which can store about 40,000 pieces of test data;

11. Built-in camera for accurate positioning;

12. PC-side software has powerful extension functions

plastic sample

Reflection measurement: adopt D/8 structure (diffuse illumination, 8° direction), conform to IS07724/1, CIE, ASTM, DIN and JIS standards.

transmission sample

Transmission measurement: adopt d:0° structure (diffuse illumination, 0° direction reception), conform to ISO, CIE, ASTM and DIN standards.

Liquid, powder, solid sample measurementMeasuring caliber

Caliber automatic identification, four calibers of φ25.4/15/8/4mm can be switched arbitrarily, taking into account the special needs of customers

Application

It is used for accurate analysis and transmission of colors in laboratories; plastic electronics, paint inks, textile and garment printing and dyeing, printing, paper products, automobiles, medical care, cosmetics and video screens, etc., in the color transmission and quality control of scientific research institutions, laboratories and other industries are widely used.


FAQ
QWhat are the requirements for setting the measurement caliber of the YS6010 colorimeter transmission mode?
A Hello, the measurement aperture/lens position in the transmission measurement mode is only in automatic mode and cannot be set manually. But make sure that the reflection measurement diameter can only be 25.4mm, and the whiteboard is not moved.
QWhat is the measurement port on the YS6010 colorimeter?
A Hello, the upper port of Weifu YS6010 is a transmission measurement port. When in transmission measurement mode, the transmission sample to be tested needs to be close to the transmission measurement port. You can use the transmission test fixture assembly to fix the sample to be tested, and close the lid above the transmission measurement port for measurement. At the same time, make sure that the reflection whiteboard is close to the reflection measurement port.
Q3Nh YS6010 colorimeter under what circumstances to black and white correction?
A Hello, the following situations need black and white correction: 1, before the first measurement after power on; 2, before the first measurement after the measurement aperture switch; 3, before the first measurement after the reflection and transmission mode switch; 4, before the first measurement after the UV mode switch; 5, when the environmental conditions change relatively large (such as temperature changes more than 5 degrees Celsius); 6, continuous long-term use of the instrument (such as more than 8 hours, the user can set their own discretion); 7, the user found that the test data is inaccurate.
QMeet GB/T7975 Desktop Colorimeter
A YS6010 Desktop Colorimeter, please refer to it.
Standard