JINGHUA UV1800PC UV-Visible Spectrophotometer Wavelength range: 190nm~ 1100nm

SE
JINGHUA UV1800PC UV-Visible Spectrophotometer with photometric, quantitative, regression equation, kinetics, multi-wavelength, DNA/protein and other functions, is a 1.7nm Light spectrum bandwidth experimental UV-Visible Spectrophotometer, UV1800PC with dedicated software and USB interface.
$2403.00/set
PromotionLoading...
Coupon
Model:

jinghua UV1800PC UV-Visible SpectrophotometerSPEC

jinghua UV1800PC UV-Visible SpectrophotometerDetails

jinghua UV1800PC UV-Visible SpectrophotometerPacking list

SKU
NB016039
Light spectrum bandwidth
1.7nm
Wavelength range
190~1100nm
Wavelength Accuracy
±0.5nm
Wavelength Repeatability
0.2nm
stray light
≤0.03%T
Noise level
0.0005A
Stability
±0.0007A/h
Baseline flatness
±0.002A
Interface
USB RS-232
monitor
Large-screen color Chinese/English touch display
Wavelength Drive
automatic
Transmittance Accuracy
≤0.3%T
Transmittance Repeatability
0.1%T
Transmittance range
0~200%T
Absorption range
-0.4~4A
Display range of concentration
0~99999
Scanning speed
High, medium and low
dynamics
have
multi-wavelength
have
DNA/protein
have
UV1800PC
UV-Visible Spectrophotometer
JINGHUA UV1800PC UV-Visible Spectrophotometer Wavelength range: 190nm~ 1100nm Picture

Shanghai Jinghua UV1800PC UV-Visible Spectrophotometer

Features

  • The application of high-performance 1200L/mm UV special grating, as well as the integration of unique monochromator optical path and imported photoelectric converter, ensure the instrument's low stray light, high resolution, and high luminosity linearity. 

  • The spacious sample chamber allows you to select a maximum 10cm sample rack according to different requirements to meet the needs of special industries. 

  • The use of imported long-life deuterium lamps and tungsten lamps ensures the instrument's high-quality requirements for light source energy, enables the instrument to be in a high-stability state for a long time, enhances reliability, and extends the service life of the instrument. 

  • It has a professional automatic wavelength error correction function to further ensure the test accuracy of the instrument. 

  • The good integration of the computer and the host, the 7-inch color Chinese/English touch screen interface, and large-scale integrated circuits make it highly intelligent and automated, reflecting the sense of technology, smooth and convenient to use, and clear and intuitive reading. Relevant data and maps of the samples being tested. 

  • The instrument has testing functions such as photometry, quantification, regression equation, kinetics, multi-wavelength, DNA/protein, peak and valley detection, etc., and can realize wavelength/time scanning, etc. Among them, the UV1800PC model has a proportional dual-beam optical path design, covering all the above-mentioned advanced function, and comes standard with an automatic eight-cell sample rack. 

  • The configuration of special software enables the instrument to have a strong memory function, which can quickly restore the test state when it is turned on again, and memorize the relevant data and spectrum curves that have been tested, thus facilitating the export of relevant data and spectrum curves. You can also import the saved relevant data parameters and spectrum curves into the worksheet for comparison and analysis with the current test spectrum.

  • The USB interface can be equipped with peripheral devices such as computers, printers, special software, keyboards or mice, and related operating tools.

FAQ
QIs there any difference between the UV1800 and the UV1800PC?
A The UV1800PC is equipped with dedicated software and a USB interface.
QWhat is the cause of spectrophotometer 100% T/0A imbalance?
A 1) Absorption value of the reference sample is too large (concentration is too high); 2) Light source lamp position shift; 3) Light source lamp is suddenly damaged when the instrument is working; 4) There is a foreign object blocking the light in the sample room; 5) Internal failure of the instrument.
QHow many samples can be set for quantitative analysis at one time?
A 8 samples.
QCan UV1800PC measure sample concentration?
A Yes, the sample concentration can be automatically tested.
Q0% T disorder, what should I do?
A The instrument needs to be adjusted to 0% T with the sample chamber cover closed. If it is not closed, it is an internal failure of the instrument and needs to be repaired.
QWavelength range of product for pentaerythritol ester measurement of this material, is there any recommended equipment?
A It is recommended that 754PC type UV1800 type UV1800PC type, UV series, UV series UV light source lamps use imported deuterium lamps from Hamamatsu, Japan, with stable performance
QWhat is the Warm-up time of the UV1800 ultraviolet spectrophotometer?
A Warm up for at least 30 minutes.
QDoes the UV-Visible Spectrophotometer meet the ISO 3856-1 specification?
A Model UV1800PC UV-Visible Spectrophotometer Compliant
Knowledge
Standard