United Kingdom

Elcometer 456 Film Thickness Gauge A456CFNFSS+T456CFNF1S

SE
External probe; with verification certificate and data management software; screen automatic swirl/spin, with tap wake-up function; can set high-low alarm; 4 Calibration methods; can set high-low alarm; Bluetooth function; reading trend chart.

Elcometer A456CFNFSS+T456CFNF1S Coating Thickness GaugeSPEC

Elcometer A456CFNFSS+T456CFNF1S Coating Thickness GaugeDetails

Elcometer A456CFNFSS+T456CFNF1S Coating Thickness GaugePacking list

SKU
NB016599
Measurement principle
Magnetic Induction + eddy Current
margin of error
±1~3%/±2.5μm
Resolution
0.1μm|1μm
Minimum measurement surface
F: 4mm;N: 6mm
Minimum radius of curvature
Convex diameter - F: 4mm N: 26mm; Concave radius - 12.5mm
probe
Split
Calibrated Certificate
Yes.
Calibrated mode
Factory calibrated, two point calibrated, zero point calibrated, single point calibrated, zero point offset, preset calibrated, Automatic calibration
Unit display
μm/mils
statistical function
Number of readings, average value, standard deviation, minimum limit, maximum limit, coefficient of variation
Interface
USB、Bluetooth
Data storage
1500 Measured values
Power Supply
2 AA batteries, rechargeable batteries can also be used
Standards
AS 2331.1.4, AS 3894.3-B, AS/NZS 1580.108.1, ASTM B 499, ASTM D 7091, ASTM E 376, ASTM G 12,BS 5599,EN 13523-1, IMO MSC.215(82), IMO MSC.244 (83)
Display screen
2.4 " (6 cm) QVGA color TFT display, 320x240 pixels
Operating temperature
80℃
Operating temperature
-10~50℃
Battery life
With 1 reading per second, it can work continuously for up to 24h
Protection level
IP64
supporting software
ElcoMaster ® Software
Machine Weight
161G (including battery)
Machine size
141*73*37mm
Measurement range
0~1500μm
Minimum headroom
88mm
A456CFNFSS+T456CFNF1S
Coating Thickness Gauge
External probe; with verification certificate and data management software; screen automatic swirl/spin, with tap wake-up function; can set high-low alarm; 4 Calibration methods; can set high-low alarm; Bluetooth function; reading trend chart.
Elcometer 456 Film Thickness Gauge A456CFNFSS+T456CFNF1S Picture


Elcometer 456 Thickness Gauge is divided into 3 types: B (basic type), S (standard type), T (advanced type), from the original Elcometer 456B type to the current 456T type. In addition, it is also divided into integral and split Thickness Gauges.

The split Thickness Gauge provides a variety of probes (probes need to be purchased separately), making measurement more maneuverable.

The British Elcometer 456 A456CFNFSS coating Thickness Gauge is a standard split magnetic induction and eddy current coating Thickness Gauge with a thickness measurement range of 0~1500 microns.

British Elcometer 456 paint Film Thickness Gauge application 1British Elcometer 456 paint Film Thickness Gauge application 2British Elcometer 456 paint Film Thickness Gauge test scene 1British Elcometer 456 paint Film Thickness Gauge test scene 2

Features

Easy to use

  • Large buttons are easy to operate even when wearing gloves

  • Multilingual menu settings

  • High contrast color LCD with auto-rotation function

  • High-low limit indicator signs

  • Factory calibrated and ready to use right out of the box

Readings are accurate

  • Measurement accuracy ±1~3%

  • Can be used in accordance with many national and international standards

  • Thermal stability measurement

  • Improves resolution readings for thin coatings

  • Accurate measurement on smooth, rough, thin and curved surfaces

Reliable measurement

  • High repeatability and reproducibility

  • 2-year console warranty

  • Full set of traceable inspection certificates provided

  • Data sets are equipped with date and time stamp functions

 solid

  • Sealed, durable and impact resistant

  • Dustproof and waterproof, protection level equivalent to IP64

  • The display is highly resistant to scratches and solvents

  • Instrument and probe construction are durable

  • Suitable for use in harsh environments

efficient

  • Fast reading: 70+/min, using ultrasound/scanning probe: 140+/min

  • Multiple calibration memory

  • alphabetic array identifier

  • User selectable calibration method

  • Compatible with ElcoMaster® and ElcoMaster® Mobile App

Powerful

  • Various probes can be used interchangeably

  • USB interface

  • Bluetooth® data output to iPhone* or Android™ device (available in standard and advanced models)

  • Can be divided into 2,500 groups to store 150,000 data

  • Capable of measuring coatings up to 31mm thick on metal substrates

Complying with formal quality system certifications such as ISO 9001 requires instruments to properly control, record and calibrate data. Users require that instrument measurement data should be traceable to various national standards. A wide range of nominal and certified diaphragms are available, along with zero plates, to ensure the accuracy of these instruments.

British Elcometer 456 A456CFNFSS coating Thickness Gauge large screen displayBritish Elcometer 456 A456CFNFSS coating Thickness Gauge screen statistics function

Elcometer 456 A456CFNFSS coating Thickness Gauge screen trend chart

Large screen display, easy to read, both metric and imperial unitsThe screen displays up to 8 statistics, and the statistical content is user selectable.On-screen trend graph shows the last 20 measured values

British Elcometer 456 A456CFNFSS coating Thickness Gauge graphic playback function

British Elcometer 456 A456CFNFSS coating Thickness Gauge can be equipped with a variety of probes

British Elcometer 456 A456CFNFSS coating Thickness Gauge side view
Playback of individual batch readings as numbers or graphs (Advanced)A variety of probes (need to be purchased separately) are available for a wide range of applications2.4" color display allows users to see data more clearly from all angles

Standard advantages and features

  • Bluetooth®; to computers, Android™ & iOSǂ devices;

  • Screen data statistics: nominal dry film thickness; NDFT, IMO PSPC standards; %>NDFT, %>90<NDFT, 90:10 test pass/test failure, high-low limit settings; sound/screen alarm can be set;

  • The number of data above the high limit;

  • The number of data below the lower limit;

  • Instant reading trend graph: data group mode;

  • Calibration methods: factory calibration (can be reset to factory calibration), zero offset (calibrated according to ISO19840 standard), automatic calibration, two-point calibration (user smooth and rough surfaces), preset calibration and measurement mode;

  • Number of data groups: 1;

  • Instrument memory: 1,500 readings;

  • Instrument (g) limit: Yes;

  • Date and time stamp: Yes;

  • Data group type: standard, calculated average, IMO PSPC;

  • Review/Purge/Delete Data Groups: Yes.

ElcoMaster data management software features

ElcoMaster® helps you review your data and generate professional reports quickly and easily. Built-in wizards guide you every step of the way, from connecting instruments to generating reports.

1. Use the ElcoMaster® software wizard to connect (Bluetooth or USB) to the Elcometer Thickness Gauge to download data quickly and easily.

2. ElcoMaster® can download the measurement results of all inspections from any Elcometer instrument.

3. Data can be stored into a simple file tree by project and detection type . The data is clearly displayed in the table

4. Using ElcoMaster®'s built-in report design, readings can be quickly displayed on images or drawings.

5. Readings from different Elcometer instruments can be printed on the same report - and quickly inserted into standard reports.

6. Suitable for cloud computing.