United Kingdom

Elcometer 456 Coating Thickness Gauge A456CNBS + straight probe

SE
Elcometer 456 Coating Thickness Gauge A456CNBS is a split Thickness Gauge for measurement of non-ferrous metal substrates, the thickness measurement range is 0-1500 μm, A456CNBS Coating Thickness Gauge has USB interface and has a variety of Calibration methods.

Elcometer A456CNBS+T456CN1S Coating Thickness GaugeSPEC

Elcometer A456CNBS+T456CN1S Coating Thickness GaugeDetails

Elcometer A456CNBS+T456CN1S Coating Thickness GaugePacking list

SKU
NB016617
Measurement principle
Eddy Current
Measurement range
0~1500μm
margin of error
±1~3%/±2.5μm
Resolution
0.1μm|1μm
Minimum measurement surface
4mm
Minimum radius of curvature
Depending on the probe selected
probe
Split
Calibrated Certificate
Yes.
Calibrated mode
Factory calibrated, two point calibrated, zero point calibrated, single point calibrated
Unit display
μm/mils
statistical function
Number of readings, average value, standard deviation, minimum limit, maximum limit, coefficient of variation
Interface
USB
Data storage
Last 5 Measured values
Power Supply
2 AA batteries, rechargeable batteries can also be used
Standards
AS 2331.1.4, AS 3894.3-B, AS/NZS 1580.108.1, ASTM B 499, ASTM D 7091, ASTM E 376, ASTM G 12,BS 5599,EN 13523-1, IMO MSC.215(82), IMO MSC.244 (83)
Display screen
2.4 " (6 cm) QVGA color TFT display, 320x240 pixels
Operating temperature
80℃
Operating temperature
-10~50℃
Battery life
With 1 reading per second, it can work continuously for up to 24h
Protection level
IP64
supporting software
ElcoMaster ® Software
Machine Weight
161G (including battery)
Machine size
141*73*37mm
A456CNBS+T456CN1S
Coating Thickness Gauge
Elcometer 456 Coating Thickness Gauge A456CNBS + straight probe Picture


Elcometer 456 Thickness Gauge is divided into 3 types: B (basic type), S (standard type), T (advanced type), from the original Elcometer 456B type to the current 456T type. In addition, it is also divided into integral and split Thickness Gauges.

The split Thickness Gauge provides a variety of probes (probes need to be purchased separately), making measurement more maneuverable.

The British Elcometer 456 A456CNBS split eddy current Thickness Gauge is a basic split eddy current coating Thickness Gauge with a thickness measurement range of 0~30mm.

Elcometer456 Eddy Current Thickness Gauge Test 1Elcometer456 Eddy Current Thickness Gauge Test 2Elcometer456 Eddy Current Thickness Gauge Application 1Elcometer456 Eddy Current Thickness Gauge Application 2

Features

Easy to use

  • Large buttons are easy to operate even when wearing gloves

  • Multilingual menu settings

  • High contrast color LCD with auto-rotation function

  • High-low limit indicator signs

  • Factory calibrated and ready to use right out of the box

Readings are accurate

  • Measurement accuracy ±1~3%

  • Can be used in accordance with many national and international standards

  • Thermal stability measurement

  • Improves resolution readings for thin coatings

  • Accurate measurement on smooth, rough, thin and curved surfaces

Reliable measurement

  • High repeatability and reproducibility

  • 2-year console warranty

  • Full set of traceable inspection certificates provided

  • Data sets are equipped with date and time stamp functions

 solid

  • Sealed, durable and impact resistant

  • Dustproof and waterproof, protection level equivalent to IP64

  • The display is highly resistant to scratches and solvents

  • Instrument and probe construction are durable

  • Suitable for use in harsh environments

efficient

  • Fast reading: 70+/min, using ultrasound/scanning probe: 140+/min

  • Multiple calibration memory

  • alphabetic array identifier

  • User selectable calibration method

  • Compatible with ElcoMaster® and ElcoMaster® Mobile App

Powerful

  • Various probes can be used interchangeably

  • USB interface

  • Bluetooth® data output to iPhone* or Android™ device (available in standard and advanced models)

  • Can be divided into 2,500 groups to store 150,000 data

  • Capable of measuring coatings up to 31mm thick on metal substrates

Complying with formal quality system certifications such as ISO 9001 requires instruments to properly control, record and calibrate data. Users require that instrument measurement data should be traceable to various national standards. A wide range of nominal and certified diaphragms are available, along with zero plates, to ensure the accuracy of these instruments.

British Elcometer 456 A456CNBS split eddy current Thickness Gauge with large screen displayBritish Elcometer 456 A456CNBS split eddy current Thickness Gauge screen statistics

British Elcometer 456 A456CNBS split eddy current Thickness Gauge screen trend chart

Large screen display, easy to read, both metric and imperial unitsThe screen displays up to 8 statistics, and the statistical content is user selectable.On-screen trend graph shows the last 20 measured values

British Elcometer 456 A456CNBS split eddy current Thickness Gauge chart playback function

British Elcometer 456 A456CNBS split eddy current Thickness Gauge can connect multiple probes

British Elcometer 456 A456CNBS split eddy current Thickness Gauge side view
Playback of individual batch readings as numbers or graphs (Advanced)A variety of probes (need to be purchased separately) are available for a wide range of applications2.4" color display allows users to see data more clearly from all angles

ElcoMaster data management software features

ElcoMaster® helps you review your data and generate professional reports quickly and easily. Built-in wizards guide you every step of the way, from connecting instruments to generating reports.

1. Use the ElcoMaster® software wizard to connect (Bluetooth or USB) to the Elcometer Thickness Gauge to download data quickly and easily.

2. ElcoMaster® can download the measurement results of all inspections from any Elcometer instrument.

3. Data can be stored into a simple file tree by project and detection type . The data is clearly displayed in the table

4. Using ElcoMaster®'s built-in report design, readings can be quickly displayed on images or drawings.

5. Readings from different Elcometer instruments can be printed on the same report - and quickly inserted into standard reports.

6. Suitable for cloud computing.

FAQ
QHow many calibration methods are there for this thickness gauge?
A Hello, there are four calibration methods: factory calibrated, two-point calibrated, zero-point calibrated, and single-point calibrated.
QDoes the ELCOMETER Coating thickness gauge meet the JJG818 standard?
A The A456CNBS + T456CN1S Coating thickness gauge can refer to
Standard