China

TIMEPOWER TP120 (k = 0.1) conductivity analyzer

Certificate

TIMEPOWER TP120(k=0.1) Conductivity analyzerSPEC

TIMEPOWER TP120(k=0.1) Conductivity analyzerDetails

TIMEPOWER TP120(k=0.1) Conductivity analyzerPacking list

SKU
NB016986
Measurement range
0.00 to 30.00 μS/cm and 0.0 to 300.0 μS/cm
Measurement error
±2.5% F.S
Resolution
0.001μS/cm
Temperature compensation range
0~60℃
Dimensions
145*120*150mm
Weight
0.64kg
Display
128X64 dot matrix LCD, Chinese display
Repeatability
≤1%
Temperature sensor
Pt1000
Water sample temperature
5~60℃
water sample pressure
<0.6 Mpa
Operating temperature
5~45℃
Operating humidity
≤ 90% RH (non-condensing)
Storage temperature
-25~55℃
Protection level
IP65
Power down save
> 10 years
Current Output
0~10mA, 0~20mA, 4~ 20mA optional, load < 300Ω
Alarm relay
2 normally open points can be set arbitrarily, AC 220V 3A/DC 30V 3A
output load Capacity
≤750Ω
Output Accuracy
1%F.S
Relay load Capacity
AC 250V 5A/DC 30V 5A
Power Supply
AC (85~ 265) V Frequency (45~ 65) Hz
Power
≤ 15W
opening size
138mm×138mm
TP120(k=0.1)
Conductivity analyzer
TIMEPOWER TP120 (k = 0.1) conductivity analyzer Picture

Introduction

TP120 conductivity analyzer is a new generation of high-end instrument with Chinese and English menu-based operation, full intelligence, multi-function, high measurement performance, strong environmental adaptability, etc., which can be used for continuous monitoring of conductivity values in solutions in various industries.

Times Xinwei TP120(k=0.1) conductivity analyzer

Features

  • The instrument adopts high-precision A/D conversion and microprocessor microprocessor technology, which can realize various functions such as conductivity measurement, temperature measurement, automatic temperature compensation, and automatic range conversion.

  • It adopts professional chip placement process and integrated high-integration circuit design.

  • The current output adopts photoelectric coupling isolation technology, which has strong anti-interference ability and can realize remote transmission. It has good electromagnetic compatibility.

  • Chinese menu-style operation, easy to understand, easy to operate.

  • Data cycle storage function, automatic removal of overflow data, easy operation, simple query, power failure data storage time of more than 10 years.

  • IP65 protection level, suitable for outdoor use.

  • The standard output signal type is optional, and the alarm relay can be set arbitrarily.

  •