◆ Humanized design
● Brushed stainless steel liner, four corners and half circle arc transition, shelves can be loaded and unloaded freely, easy to clean the studio.
● Water-proof heating method ensures that the temperature of the studio is uniform, and the constant temperature can still be maintained for a long time in the power-off state.
● There is a glass door in the box that is easy to observe, and in the working state, the user can open the glass door at any time, and the Incubator will be automatically suspendedwork, and cut off the heating with the circulating fan.
● Microcomputer controller; Self-tuning PID technology, no temperature overshoot; It has temperature, water level and door opening alarm, etcThe user-friendly design function can avoid the destruction of experimental samples by sudden temperature changes.
◆ Automatic control of circulating fan speed
● The speed of the circulating fan can be automatically controlled, and when the temperature in the box is in a constant state, the speed will be reduced, and the circulating air speedThe wind speed will be adjusted to the appropriate air speed for cell growth to avoid volatilization of the sample due to excessive air volume during the test.
◆ LCD multi-segment programmable controller(with "N"With this feature)
● Multi-segment LCD programmable controller, one-screen display of various parameters, menu-based operation interface, can simplify complex testsprocess, truly realize automatic control and operation.
● Parameters such as multi-stage temperature, circulating wind speed, time and heating rate can be set and programmed at the same time, and the temperature can be raisedGradient control, slow heating up from the initial temperature in the box, etc., can also be preset to automatically start, standby and shutdown functions.
◆ Safety features
● Independent temperature limit alarm system, sound and light alarm to prompt the operator, to ensure the safe operation of the experiment without accidents.
● It has alarms such as low temperature, high temperature, ultra-temperature and water shortage.
● It has an alarm for too long to open the door.
◆ Ultraviolet sterilization system (Matching)
● The inside of the chamber can be disinfected regularly, which can effectively kill the floating bacteria in the circulating air in the chamber, so as to effectively prevent pollution during cell culture.

Options:(Option lead time increased by 7 days)
1. Multi-segment programmable liquid crystal temperature controller
2. Embedded printer
3. Independent temperature limiting controller
4. RS485 interface and communication software
5. Ultraviolet sterilization system
6. Automatic water filling system is optional
Note: RS485/232, printer can choose one
Model Technical Indicators | GHP-9050 GHP-9050N | GHP-9080 GHP-9080N | GHP-9160 GHP-9160N | GHP-9270 GHP-9270N |
Supply Voltage | AC220V 50HZ | |||
Heating method | Water-jacketed | |||
Temperature control range | RT+5~65℃ | |||
Temperature resolution/fluctuation | 0.1℃/±0.3℃ | |||
Temperature uniformity | ±0.5°C (37°C at test point) | |||
Operating ambient temperature | +5~35℃ | |||
Input power | 450W | 650W | 850W | 1350W |
volume | 50L | 80L | 160L | 270L |
Inner tank size (mm): W×D×H | 350×350×410 | 400×400×500 | 500×500×650 | 600×600×750 |
Dimensions (mm): W×D×H | 500×500×700 | 550×550×800 | 650×640×950 | 750×740×1050 |
Cargo carrier (standard) | 2 pieces | |||
Timing range | 0~9999min | |||
Note:"N"It is a multi-segment program LCD controller