China

JINGTIAN NDJ-1S digital display Rotational Viscometer 100,000 CP

SE
JINGTIAN NDJ-1S digital display Rotational Viscometer has a Viscosity range of 20~100000mPa·s. It is a liquid crystal display Viscometer. NDJ-1S digital display Rotational Viscometer is equipped with L1, L2, L3, L4 rotors.
Model:
Certificate

JingTian NDJ-1S Digital Rotational ViscometerSPEC

JingTian NDJ-1S Digital Rotational ViscometerDetails

JingTian NDJ-1S Digital Rotational ViscometerPacking list

SKU
NB020480
Display
digital display
Measuring range
20~100000mPa·s
Rotor Specifications
L1#、L2#、L3#、L4#
Reproducibility
±1%F.S
Measurement Accuracy
±5%
Power
Dedicated power adapter (input 100~240V 50~60Hz)
Rotor speed
6/12/30/60rpm
Display information
Viscosity, rotational speed, percentile torque, rotor number and the maximum measurable Viscosity value of the selected rotor at the current rotational speed, etc
NDJ-1S
Digital Rotational Viscometer
JINGTIAN NDJ-1S digital display Rotational Viscometer 100,000 CP Picture

Introduction

The LCD of Shanghai JINGTIAN NDJ-4S digital display rotary viscometer can display viscosity, shear rate, shear stress, rotational speed, temperature, percentile torque, rotor number and the maximum measurable value of the selected rotor at the current rotational speed. Viscosity value etc. It is a digital display rotary viscometer with a measuring range of 20~100000 mPa·s.




FAQ
QOur application requires high accuracy viscosity measurement. Is the NDJ-1S suitable?
A The measurement accuracy of NDJ-1S is +/- 5% (FS), which is more suitable for general medium and low viscosity measurement. If your application requires higher accuracy, you can consider models with higher accuracy.
QThe laboratory often measures medium and low viscosity oils and Coatings. Can NDJ-1S meet our needs?
A Yes. The measurement range of NDJ-1S is 2~ 100000cp, which is very suitable for the measurement of medium and low viscosity samples such as oils and coatings.
QCan 1s digital rotational viscometer show shear stress?
A Hello, yes.
QAre all Viscometers available for inspection in the JJG1002 standard?
A Not necessarily, but this NDJ-1S Viscometer can be considered.
QWhere is the temperature probe interface of Jingtian NDJ-1S digital display rotational viscometer?
A Hello, the Viscometer temperature probe interface of Jingtian ndj-1s model is on the upper part behind the Viscometer host, which is also next door to the RS232 interface.
QWhat is the function of the reset button on the Viscometer of Jingtian NDJ-1S digital display?
A Hello, the reset button can restart the application and stop the motor from working, and the host enters the standby state.
QI haven't done any testing before, and I don't know how to do it. How to choose a Viscometer?
A It is easy to make the results of the viscosity test. After the measurement, he said that it is different from others, so it is wrangling. If the next home has viscosity requirements for his products, ask the next home for detection methods. If the purchase inspection, ask the previous home for Measurement conditions and methods. There are many viscosity measurement methods, and the measurement results cannot be converted.