China

JINGTIAN RVDV-2 digital display Rotational Viscometer 40 million mPa·s 200rpm

SE
Fine digital display Rotational Viscometer RVDV-2 measurement Viscosity range 40~40000000mPa·s, is a liquid crystal display Viscosity Measurement Instrument, used for Coating paint and other Viscosity tests, RVDV-2 measurement accuracy of ± 1%.
Model:
Certificate

JingTian RVDV-2 Digital Rotational ViscometerSPEC

JingTian RVDV-2 Digital Rotational ViscometerDetails

JingTian RVDV-2 Digital Rotational ViscometerPacking list

SKU
NB020600
Display
digital display
Measuring range
40~40000000mPa·s
Rotor Specifications
R2#、 R3#、R4#、R5#、R6#、R7#
Reproducibility
±0.5%F.S
Measurement Accuracy
±1%
Power
Dedicated power adapter (input 100~240V 50~60Hz)
Rotor speed
0.1~200rpm
Display information
Viscosity, shear rate, shear stress, rotational speed, temperature, percentile torque, rotor number and the maximum Viscosity value that can be measured at the current rotational speed of the selected rotor, etc
Speed control method
CVT
RVDV-2
Digital Rotational Viscometer
JINGTIAN RVDV-2 digital display Rotational Viscometer 40 million mPa·s 200rpm Picture

Introduction

The LCD of Shanghai JINGTIAN RVDV-2 digital display rotary viscometer can display viscosity, shear rate, shear stress, rotational speed, temperature, percentile torque, rotor number and the maximum measurable value of the selected rotor at the current rotational speed. Viscosity value etc. It is a digital display Rotational Viscometer with a range of 40 to 40,000,000 mPa·s.

Shanghai JINGTIAN RVDV-2 digital display rotary viscometer

Dimensions

Dimensions of digital display viscometer 2Dimensions of digital viscometer 1

FAQ
Q10 1/S, viscosity between 9000-16800mPa.s
A RVDV-2, R7 rotor, the shear rate at 47.8RPM is 10 1/S and then matched with Thermostatic bath.
QIf it is low viscosity, this one uses L0 or R1 rotor. Is R1 standard?
A Rotors R1 and L0 are optional and can be used
QRVDV-2, I would like to ask, what is the viscosity range of the No. 2 rotor standard on this machine?
A R2 series rotor viscosity range is 40-400,000 mPa
QTo use cream for testing, is there a commonly used medicinal viscometer?
A This RVDV-2 is enough.
QCan this one measure xanthan gum?
A Yes, there are practical application cases.
QCan this Viscometer display shear rate and shear force?
A OK
QMeasure gel viscosity, but the sample volume is not large, you need to use a small sample adapter, the detection range is less than 100,000, and the temperature is about 37 degrees. Viscometer recommends
A RVDV-1 Viscometer, plus a small amount of sample adapter, plus super Thermostatic bath can be
QIs there any application difference between RVDV and HBDV? It is used on colloids.
A RVDV can measure sealant, silicone, nail-free glue, etc., HBDV can measure silicone rubber, Glass glue and other more viscous
QCan the rotor be tested directly after installing a digital rotational viscometer?
A Hello, after the rotor is installed, it should wait for ten minutes after being immersed in the tested liquid, so that the temperatures of the two are consistent before measurement
QWhich Viscometer displays shear rate and shear force?
A The DV-2 series has shear rate and shear force display. The DV programmable download software can program the shear rate change of the Viscometer in advance, and then collect and draw data through the data collection software