China

JINGTIAN, Small Sample Adapter, Rotor

SE
JINGTIAN, a small number of sample adapters are equipped with one rotor of No. 21, 27, 28, and 29 each. The sample volume is small, only 10~20mL is required. It is especially suitable for valuable sample testing. It is suitable for NDJ-5S, NDJ-8S, SNB-1, SNB-2, SNB-3, SNB-AI, LVDV-1, RVDV-1, HBDV-1, LVDV-2, RVDV-2, HBDV-2 Viscometer of the same brand.
Model:
Certificate

JingTian rotor Small sample adapterSPEC

JingTian rotor Small sample adapterDetails

JingTian rotor Small sample adapterPacking list

SKU
NB020629
rotor
Small sample adapter
JINGTIAN, Small Sample Adapter, Rotor Picture

Introduction

[Adapter for Small Samples]

. With 21, 27, 28, 29 rotors each one

. Small sample volume, only 10-20mL, especially suitable for expensive sample testing

. A constant temperature bath connected with external circulation function can control the temperature of the tested sample

. Easy to disassemble and clean

Shanghai JINGTIAN Small Sample Adapter Rotor

FAQ
QIs there a probe to detect the temperature of the sample in that adapter? That is, how do I know that my sample has dropped to 25 degrees Celsius during the cooling process?
A Use the RTD temperature sensor to display the sample or Environmental temperature directly on the Viscometer display
QMy SNB-1 now only has rotors 0-4 to choose from. When choosing rotors, I can't choose No. 21, 27, 28, or 29. After I add a small number of sample adapters, can this Viscometer choose these rotors?
A Yes, only after a small amount of adapters can you use/select the 21, 27, 28, and 29 rotors. 0-4 is the standard rotor, more than 300ml, so the machine can only set these 4 rotors. After a small amount of adapters, you can use the small rotor.
QCan the No. 3 and No. 4 rotors use this adapter?
A A small number of sample adapters are equipped with the four rotors of No. 21, 27, 28, and 29, because between the inner diameter of the container and the diameter of the drill, it represents a shear coefficient, which is fixed and used together. The 3 and 4 rotors are all in our regular container, and then a u-shaped frame is added to the regular beaker, which is the standard test. But the No. 3 rotor is not good, it has an impact. As for the No. 4 rotor, for those with relatively large viscosity, it seems that there is no flowability. So for the sample in this case, the No. 4 rotor puts the sample in the small amount of sample adapter, and then takes the No. 4 rotor inside to test, which may not have a great impact, because his sample will not spread out, and the size of the container does not have a great impact on it
QIs this small sample adapter tested with the original Viscometer rotor, or is there a separate rotor?
A This one is equipped with one rotor each of No. 21, 27, 28, and 29.
QHow much does this weigh?
A About 2kg
QCan a small number of sample adapters be used on the snb-2? How many rotors can be used, and how many milliliters of sample can be used?
A A small number of sample adapters are equipped with one rotor of No. 21, 27, 28, and 29 as standard, and the sample volume is less than 10~ 20mL, which can be used in snb-2.
QFind a small sample adapter for JB/T9357
A This rotor small sample adapter is.
Standard