HANNA BL121-10 Water quality measurement controller Specifications

Search
HANNA BL121-20 Water Quality Measurement Controller

Brand:HANNA

Model:BL121-10

Name:Water quality measurement controller

DetailsContact
Specifications
PH range 0.00~14.00pH
pH resolution 0.01pH
Measurement Accuracy of pH ±0.05pH
pH calibrating mode Acidity measurement calibrate: Manual calibration of single point; acidity pH buffer calibrate: two points automatic identification calibrate, built-in standard points: 4.01, 7.01, 10.01 pH
pH temperature compensation automatic
ORP Measurement range ±2000mV
ORP resolution 1mV
ORP Accuracy ±5mV
ORP calibrated mode Manual calibration
Measurement range -5.0~105.0℃
Temperature Resolution 0.1℃
Measurement Accuracy ±1℃
Dosing settings pH, ORP: Adjustable set point and proportional band; delay is proportional to enable startup at power-up and over-protection
alarm settings When 5 consecutive readings exceed/fall below the set value, trigger [start] all parameters to enable/disable the selected LED alarm light, LED intuitive alarm system based on the alarm filtrate option, the user sets the alarm relay control item
Alarm relay output [1] SPDT 5A/230 VAC activated by pH/ORP/temperature optional alarm conditions
analog output 4~ 20 mA output configuration; Output Impedance ≤ 500 ohms Accuracy < 0.5% FS, up to 50 V relative to isolation current
Auxiliary input [3 pieces] Low level in acid/alkali tanks (open contact) Low level in chlorine tanks (touchpoint on); hold input (touchpoint on)
Pump control settings Automatic and manual modes; Flow rate adjustable from 0.5 to 3.5 L/h
data management All data.csv files transferred via USB flash drive; automatic recording of pH, ORP and temperature measurements, GLP and events including alarms, errors and power failures, data stored for 60 days, Data Logging interval of 10 seconds
password protection Settings, calibrate and log call option features are all password protected by user settings
GLP function Automatic recording of acidity pH/oxidation reduction ORP electrode calibrated, date and time information
Power Supply AC100-240V, Power 10VA
Applicable environment 0~50 ℃; max95% RH non-condensing
Dimensions 245*188*55mm
Weight 1700 g
pH temperature compensation range -5.0~105.0℃
Packing list
machine X1, manual X1, certificate X1, warranty card X1
After-sales instructions

Warranty days: 365 days

[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/20698/Specification.html , which is only used by users who purchase commodities from the NBCHAO.