1. Measurement accuracy: ± 10%
2. Resolution: 0.1 uW/cm2
3. Spectral response: center point 254nm, bandwidth 230nm - 280nm
4. Measuring range: 0-8000 uW/cm2
5. Host size: length 130mm * width 70mm * height 28mm
6. Probe size: diameter 45mm * height 16mm
7. Probe cable length: 1 meter
8. Instrument weight: about 250 grams
9. Batteries, 4 AAA alkaline dry batteries
1. Suitable for short-wave sterilization ultraviolet intensity measurement
2. 254nm wavelength ultraviolet measurement
3. Dedicated UVC probe with correction filter to eliminate the influence of UVA and UVB
4. Data memory and viewing of maximum value and real-time value

,
According to WS/T 367-2012 "Technical Specifications for Disinfection of Medical Institutions", the measurement is to ensure that the UVC ultraviolet radiatIon Meter is located at the center of 1 meter below the lamp tube, and the new lamp tube is qualified if it is greater than 90uw/cm2, and the old lamp tube in use is lower than 70uw/cm2 is unqualified. Our company also produces a UV ruler hook for purchase to facilitate on-site UV intensity measurement.