Germany

Germany IKA HRC 2 controlled heating refrigeration thermostatic circulator 4L

Germany IKA HRC 2 control heating refrigeration thermostatic circulator is a temperature control of -30~100 ℃ circulating Water Bath, bath is stainless steel, HRC 2 control has a tank capacity of 1.4~ 4L, with sound and light alarm function.

IKA HRC 2 control ThermostatSPEC

IKA HRC 2 control ThermostatDetails

IKA HRC 2 control ThermostatPacking list

SKU
NB021544
Temperature range
-30~100℃
Temperature sensor
Pt100
Temperature sensor
Pt100
Pump type
Pressure/suction pump
Flow rate of pump
Adjustable
Power supply Voltage
230 / 115 V
Dimensions
220*475*525mm
Net Weight of the whole machine
33.3kg
HRC 2 control
Thermostat
Germany IKA HRC 2 controlled heating refrigeration thermostatic circulator 4L Picture

Introduction

HRC2 control is an energy-efficient, compact heating, refrigeration thermostat with an operating temperature range of -30°C to 100°C, adjustable pressure suction binary pump, and inverter compressor to reduce noise and operating costs. 

The HRC2 in combination with an external temperature sensor enables precise temperature control in closed or open circulation applications. 

The 4-liter insulated bath allows for fast heating and reduces the effects of heat transfer at low temperatures. For external circulation applications, the usable volume is up to 2.5L. 

The funnel dosing and drain design allows for easy and safe replacement of the heat transfer medium. The air filter for cleaning the compressor can be easily removed without any tools. The drain port at the front of the bath is designed to be safer, and the hose can be connected and the drain valve can be opened to drain the bath, so that the user will not have any contact with the heat transfer medium.

Product configuration

Inverter compressor, refrigeration power 400W

Circulating pump made of PEEK: (0.5 bar; 22 l/min)

Heating power 1500 W (1200 W @115V)

Liquid level monitoring display

Safety class III (FL) according to DIN 12876

Adjustable safety loop


HRC2 control comes standard with a Wico wireless controller, which can be operated remotely, and has a programmed temperature control function, 10 programs, 10 steps per program. 

Extra features for control:

D mode (confirmation mode)

A signal that the set point has been reached

Timing/Timing

Exhaust function when the reactor is connected


interface

Interface for external PT100 temperature sensor (option: PT100.30, Lemo interface)

Interface for connecting solenoid valves (Mulit IO)

Alarm interface (Multi IO)

Standby input interface (Multi IO)

RS 232

USB