China

KSJ MG26-F2 intelligent Gloss meter projection angle 20 °/60 °

SE
20 °/60 ° dual angle measurement, including haze and reflectance measurement, MG26-F2 boot self-calibration and self-diagnosis, 10,000 measurement data storage capacity, support for deleting data, standard Gloss data processing software.
Model:
Certificate

KSJ MG26-F2 Gloss meterSPEC

KSJ MG26-F2 Gloss meterDetails

KSJ MG26-F2 Gloss meterPacking list

SKU
NB002160
Geometry
20°/60°
Measuring range
Gs(20°):0~119.9、Gs(60°):0~119.9;Gs(20°):120~2000 Gs、(60°):120~1000
Repeatability
0.2GU;0.2%
Reproducibility
0.5GU|0.5%
Measuring Areas
9*10mm(20°);9*16mm(60°)
Error of indication
±1.5; ±1.5%
Division Value
0.1GU|1GU
Standards
ASTM D523、GB/T13891、GB7706、GB8807、JJG 696-2002
Dimensions
155*48*75mm
Weight
400g
Power
One AA (or LR6) alkalinity dry or rechargeable battery
Measurement port size
12*28mm
Zero Error
0.2GU
Operating Temperature
10~40℃
Storage temperature
-10℃~60℃
Relative Humidity
< 85% no condensation
Operating Time
One alkalinity dry battery can support more than 20,000 measurements
MG26-F2
Gloss meter
20 °/60 ° dual angle measurement, including haze and reflectance measurement, MG26-F2 boot self-calibration and self-diagnosis, 10,000 measurement data storage capacity, support for deleting data, standard Gloss data processing software.
KSJ MG26-F2 intelligent Gloss meter projection angle 20 °/60 ° Picture

product description

KSJ MG26-F2 intelligent gloss meter is a gloss measuring instrument designed and manufactured according to the international standard ISO2813 and Chinese national standard GB/T9754. The Specifications of the intelligent gloss meter meet the requirements of ASTM D523, GB/T13891, GB7706, and GB8807 standards. All performance indicators meet the requirements of the first-class working machine in the national JJG 696-2002 "Specular Gloss Meter and Gloss Board" metrological verification regulations.

The gloss value of the intelligent gloss meter can be traced back to the national glossiness benchmark of Shanghai Institute of Metrology and Testing Technology (SIMT).

The main function

  • Multi-angle one-key to complete the measurement

  • Optional angle mode

  • Power-on self-diagnosis, self-calibration

  • Practical and rich display interface: Simultaneously display each current measurement value, average value, group sequence number and measurement sequence number

  • The measurement data is pre-grouped, and each measurement group can contain a maximum of 99 measurements

  • The unit of measurement defaults to Gloss Units (GU), and can also be switched to Percent Reflectance (%)

  • Grouped measurement data can be browsed by specifying the group number

  • Immediate deletion of invalid measurements

  • Comes with USB adapter and Glossy data processing software

  • It has the functions of low power prompt and full storage prompt

  • With haze value calculation and percentage reflectance unit conversion function

Application

KSJ MG26-F2 intelligent gloss meter is widely used in many fields such as paint coatings, decoration materials, building materials, plastic materials, bamboo and wood products, ceramic products, leather products, film paper, printing ink, automobile maintenance, molding molds, etc. Gloss measurement of material and article surfaces.

structure diagram

KSJ MG26-F2 intelligent gloss meter front structure diagram

Back structure diagram of KSJ MG26-F2 intelligent gloss meter

Structural diagram of KSJ MG26-F2 intelligent gloss meter working board

Structural diagram of the battery compartment of KSJ MG26-F2 intelligent gloss meter

Structural diagram of KSJ MG26-F2 intelligent gloss meter calibration work board

plan

KSJ MG26-F2 Intelligent Gloss Meter Display

FAQ
QCan KSJ's glossmeter measure the average?
A Yes, all four models of the F2 Smart Series can be automatically averaged.
QHow many times can this glossmeter measure data?
A Hello, a dry battery can test data more than 20,000 times.
QWhat is the Gloss meter calibrated of Keshijia MG26-F2? Is it automatic?
A Hello, the calibrate of this gloss meter is automatically inspected and calibrated when it is turned on, or you can calibrate it manually.
QWill the strong light environment affect the measurement of glossmeters? How to solve?
A Strong light environment will affect the measurement of glossmeter, it is recommended to avoid strong light environment measurement, if you want to avoid external light leakage, you can use black flannel around the measuring port to block the rays of light.
QWhat is the difference between WGG60 series and MG series Glossiness?
A WGG60 series host power supply is built-in rechargeable lithium battery. MG series host power supply is an AA (No. 5) rechargeable or alkaline battery. If the use Frequency is high, use WGG60 series. On the contrary, if the use Frequency is low, use MG series, but please take out the battery after use to protect the instrument host.
QAre all intelligent glossmeters available for detection in the ASTMD3794 standard?
A Not necessarily, but this MG26-F2 intelligent glossmeter can be considered.
QCan the MG26-F2 Gloss meter recover after deleting the data by mistake?
A Hello, no, the data of this MG26-F2 Gloss meter measurement cannot be recovered after emptying, it is recommended to be careful when using it.
QWhat is the Geometry of this MG26-F2 Gloss meter? Is there any data software?
A Hello, MG26-F2 Gloss meter is a dual angle measurement, 20 ° and 60 °. This Gloss meter comes with Gloss data processing software.
Knowledge
Standard