Germany

TESTO testo 175 T3 temperature Recorder

The TESTO testo 175 T3 temperature Recorder with 2 external Thermocouple probe interfaces (suitable for connecting T-type and K-type Thermocouples) can simultaneously record the temperature Measured value of 2 channels.

testo testo 175 T3 Temperature RecorderSPEC

testo testo 175 T3 Temperature RecorderDetails

testo testo 175 T3 Temperature RecorderPacking list

SKU
NB025788
Type K (NiCr-Ni) Measurement range
-50 ~ +1000 ℃
Type K (NiCr-Ni) Measurement Accuracy
+/- 0.5 ℃ (-50~ + 70 ℃) +/- 1 Digit +/- 0.7% Measured value (+ 70.1~ + 1 0 0 0 ℃) +/- 1 Digit
Type K (NiCr-Ni) resolution
0.1 ℃
Type T (Cu-CuNi) Measurement range
-50 ~ +400 ℃
Type T (Cu-CuNi) resolution
0.1 ℃
Weight
440 g
diameter
95*55*30 mm
Environmental temperature
-20~+55℃
Protection level
IP65
Number of channels
2 channels, external
Shell material
Plastic
standard
EU Directive 2014/30/EU; 2011/65/EU
Measurement ratio
10 s - 24 h
Power Supply
3 x AIMn type AAA or Energizer
Battery life
3 years (measurement period is 15 minutes, + 25 ℃)
Data storage
1,000,000 Measured value
Storage temperature
-20~+55℃
testo 175 T3
Temperature Recorder
TESTO testo 175 T3 temperature Recorder Picture

Description

The testo 175 T3 dual-channel temperature logger with 2 external thermocouple probe connections (for connecting thermocouples type T and K) enables simultaneous recording of temperature measurements in 2 channels. For example, when checking whether a heating system is working properly, a two-channel temperature data logger measures the inlet and return air temperatures at the same time. At the same time, the temperature logger can also be used in a variety of industrial production control links where the temperature values of different measuring points need to be monitored at the same time.

Cutting-edge technology that keeps pace with the world and provides a high level of data security

We have a large number of external thermocouple temperature probes that you can choose from to work with your temperature loggers. Customers can freely choose the corresponding probe according to the application requirements, including surface temperature probe, air temperature probe and immersion/penetration temperature probe. The corresponding measuring range and measurement accuracy depend on the selected probe. In principle, a temperature range of -50 °C to +1,000 °C is sufficient with the appropriate probe.

The data logger's display displays a large amount of important data, including current readings, set limit values, overrun points, maximum/minimum values, and remaining battery level. Since all this data can be displayed directly through the display, it can be read directly without having to turn on a computer.

In addition, the data logger is equipped with a large internal memory that can store up to 1,000,000 sets of data and a battery life of up to 3 years. These product features give you the freedom to choose when to read the data in the logger without having to read it frequently. Powered by standard AAA batteries, user-replaceable. Data loggers with up to IP 65 protection ensure that the device is protected against dust and water when used in high-dust environments.

The testo 175 T3 temperature data logger offers a high level of data security and integrity, and data stored in the memory is not lost even when the battery runs out or is replaced.

Programming and data reading of the temperature logger

The new generation of Testo humidity and temperature loggers supports the following three software versions to meet the customer's needs for instrument setup, data reading and analysis:

The ComSoft Basic software – free to download – provides quick and easy instrument programming and data reading

ComSoft Professional – optional – provides more professional data analysis and management functions

ComSoft CFR 21 Part 11 Industry Specialty Software – Optional – Ideal for Industry and 21 CFR Part 11 Compliant

Please note: The instrument needs to be connected to at least 1 temperature probe (optional) for normal operation. In addition, you will need a USB cable (not included in the product package) to connect to a computer for instrument programming. Data can be read and transferred to a PC via a USB cable or SD card (optional).

FAQ
QIs the price listed on the website the final cost for overseas purchases?
ANo, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
QWhat is the specific process for purchasing the product?
AFirst, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
QWhat should I do if the instrument I purchase needs repair?
AAfter-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.
Knowledge
Standard