China

Qiwei TND-4T Benchtop Coating-4 Viscometer Pure Copper/4mm

SE
Comply with GB/T 1723 standard; brass material with bracket; aperture 4mm; capacity 100ml.
Model:
Certificate

QiWei TND-4T Tu -4 ViscometerSPEC

QiWei TND-4T Tu -4 ViscometerDetails

QiWei TND-4T Tu -4 ViscometerPacking list

SKU
NB025827
outflow time
0~150s
Flow hole diameter
4mm
Capacity
100ml
Cup body material
pure copper
Applicable standard
GB/T 1723
Correction factor
K=1±0.03
TND-4T
Tu -4 Viscometer
Comply with GB/T 1723 standard; brass material with bracket; aperture 4mm; capacity 100ml.
Qiwei TND-4T Benchtop Coating-4 Viscometer Pure Copper/4mm Picture

Principle and Calculation

The viscosity measured by the Tu-4 viscometer is the conditional viscosity, which is the time for a certain amount of sample to flow out of a hole with a specified aperture at a certain temperature, expressed in seconds. Convert the obtained sample outflow time (s) into kinematic viscosity value (㎜²/s) with the following formula:

When t<23s, t=0.154μ+11

When 23s≤t≤150s, t=0.223μ+6.0

In the formula:

    t——outflow time, s;

    μ—kinematic viscosity, mm²/s.

main technical characteristics

1. The Tu-4 viscometer is made of brass, with a capacity of about 100ml. Its geometric dimensions and inner wall roughness meet the requirements of "GB/T 1723 Method for Determination of Viscosity of Paint". The bottom is a stainless steel outflow hole.

2. Correction coefficient K: Multiply the viscosity measured by the calibrated viscometer by the K value, which is the conditional viscosity measured by the viscometer. The correction coefficient K should be within the range of 1±0.03, otherwise the viscometer should be replaced. The viscometer should be calibrated periodically according to the frequency of its use.

3. The difference between the measured value and the calculated value of the outflow time of the standard viscosity liquid with known kinematic viscosity from the Tu-4 viscometer should be within ±3% of the calculated value.

FAQ
Q Why does the correction factor K of the Tu-4 Viscometer need to be regularly corrected?
A Measurement accuracy of Viscometer may vary due to Frequency of use and external factors. Correction factor K should be within the range of 1 +/- 0.03, otherwise Viscometer needs to be replaced to ensure the accuracy of measurement results.
QHow to Use Desktop Paint-4 Viscometer?
A Pour 100 ml of the liquid to be tested into the viscosity cup, let the liquid flow out from the 4 mm flow hole at the bottom at a certain temperature, measure the flow time, and then convert the flow time into a kinematic viscosity value according to the formula.
QWhat should I do if the correction coefficient K of the viscosity cup is out of range?
A If the correction factor K is outside the range of 1 +/- 0.03, the Viscometer should be discontinued and corrected. If it cannot be corrected, the Viscometer needs to be replaced.
QDoes this 4 viscosity cup come standard with the following receiving cup?
A There is.
QDoes the Tu-4 Viscometer meet the specifications of JT/T821.4?
A Model TND-4T Coated -4 Viscometer Compliant
QWhat is the outflow diameter of the Pushen T-4 viscosity cup in mm?
A Hello, 4MM.
QCan the viscosity of nitro paint be measured by coating 4?
A OK
Knowledge
Standard