Full-color large-screen LCD display and screen touch operation;
Four pre-set fixation programs (package, solution, instrument, test) and two custom spaces for users to personalize their work programs;
The control system with microcomputer modular design, dual PID temperature control of Water Tank and sterilization room;
Configure the key parameters of the microprinter print sterilization process;
Dual pressure, double temperature, digital dynamic display and pressure gauge display;
The screen adopts the method of man-machine dialogue and prompts to effectively avoid the occurrence of wrong operation;
The door opening and locking are electrically electric, and the control is flexible and reliable;
The sterilizer is designed with its own steam generator and jacket structure;
The sterilization chamber and jacket are respectively equipped with overVoltage protection devices;
The jacket and sterilization chamber are respectively controlled by over-temperature alarm, and the sterilization procedure is stopped after the alarm;
The sterilization conditions (jacket preheating temperature, door locking and sealing, etc.) do not meet the requirements and cannot enter the sterilization procedure;
The sterilizer is equipped with a safe exit procedure, which interrupts the sterilization procedure for any reason, and after confirming the exit, the system automatically ends the sterilization work according to the "safe exit" method;
If there is a water shortage during the working process of the sterilizer, the heating power supply can be automatically cut off and the water supply can be replenished, and the water replenishment will continue to work after the water replenishment;
At the end of the sterilization work, after the temperature and pressure reach the door opening condition, the alarm and the screen will prompt at the same time;
Integrated water ring Vacuum Pump, low noise and vacuum degree can reach -0.095MPa or less;
The maximum sterilization temperature can reach 135°C.