Rongjida DM-III minimum film formation temperature Tester

Rongjida minimum film formation temperature meter DM-III layer interval of 20mm, layer plate temperature range -10 ℃~ 50 ℃ between arbitrary adjustment, DM-III layer plate detection point 13 points, inspection table shows 16 points.
Model:
Certificate

RongJiDa DM-III Minimum film formation temperature TesterSPEC

RongJiDa DM-III Minimum film formation temperature TesterDetails

RongJiDa DM-III Minimum film formation temperature TesterPacking list

SKU
NB027328
Layer operation temperature range
-10~50℃
Layer spacing
20mm
Layer board detection point
13 point
Inspection form display
16 point, of which 1 to 13 point is the working layer temperature, 14 point is the instrument environment temperature, 15 and 16 point are the inlet and outlet temperature of the chilled down water
DM-III
Minimum film formation temperature Tester
Rongjida DM-III minimum film formation temperature Tester Picture

Introduction

Developed according to GB/T9267-2008 standard. At the same time, it meets the test requirements of GB/T 11175-2002. It is used to determine the minimum film-forming temperature of polymer emulsions and interface treatment agents. When polymer emulsions are used as surface treatment agents for coatings, adhesives, chemical fiber fabrics, leather and paper, their film-forming properties are one of the most important properties. . Using multi-channel inspection digital display, the temperature control data is accurate, and the calibration is convenient. Fully automatic measurement, no human error, simple, accurate, fast and convenient. It can simply, intuitively and accurately determine the minimum film-forming temperature of polymer emulsion.

Product principle

A cooling source and a heating source are respectively set on a suitable metal plate, and kept at the constant temperature point of the set temperature, and different temperature gradients are generated on the metal plate through the principle of metal heat conduction. A uniform thickness of sample is coated on the temperature gradient plate, and the sample is subjected to the temperature on the gradient plate to evaporate water to form a film. Due to the different temperatures on the gradient plate, the phenomenon of film formation on the sample is also different. The part where the dividing line is found is the MFT temperature of the sample.


FAQ
QHow to use a minimum film forming temperature (MFFT) meter?
A Using a minimum film forming temperature (MFFT) meter requires a series of steps, including preparing the instrument, setting up cooling and heating presets, preheating the instrument, Spreader Coating, and observing the formation of film formation boundaries.
QDoes the RONGJIDA minimum film forming temperature (MFFT) meter meet the ISO 2115 standard?
A The DM-III minimum film forming temperature (MFFT) meter can be referred to