| Accuracy class | Ⅰ |
| Maximum weighing | 100g |
| Minimum weighing value | 10mg |
| Actual Graduation d | 0.0001g |
| Verification Graduation e | 0.001g |
| Repeatability Error | ±0.0003g |
| MPE | The load m is represented by the verification Graduation e |
| ±0.5e | 0≤m≤50000 |
| ±1.0e | 50000<m≤200000 |
| ±1.5e | 200000<m |
| Settling time | ≤8s |
| Operating Temperature | 17.5 ℃~ 22.5 ℃, Temperature Fluctuation is not more than 1 ℃/h |
| Relative Humidity | 50% RH~75% RH |
| Pan size | Ф80mm |
| Wind Shield size | 220*200*250mm |
| Dimensions | 350*220*350mm |
| Gross Weight | 9kg |
| Power | AC adapter 220V 50Hz output DC9V-2.2A |
Warranty days: 365 days
1. For product quality problems caused by non-human products of instruments and equipment, the manufacturer will provide free warranty services during the warranty period;
2. There is no warranty period for fragile and consumable parts, and if you find product quality problems after receiving the goods, please apply for return and exchange within 7 days, and you will not wait for the overdue ones;