INESA WBL-830 Automatic specific surface area and porosity analyzer Specifications

Search
INESA INESA WBL-830 Automatic specific surface area and porosity analyzer

Brand:INESA

Model:WBL-830

Name:Automatic specific surface area and porosity analyzer

DetailsContact
Specifications
specific surface area 0.01 m2/g to unlimited
Analysis range 3.5 to 5000 Angstroms
test principle Low temperature nitrogen physical adsorption (static Volumetric method)
Adsorption gas nitrogen
P/P0 range 1×10-6―0.995
Pressure and temperature measurement Imported absolute pressure sensor 0-133KPa, Accuracy 0.12%, 6 temperature sensors: PT-100, Accuracy 0.1 ℃ 1
Dewar 2L, duration 80 hours
Vacuum pump mechanical pump
Ultimate Vacuum 1.0×10-4Torr
Measurement software Adsorption/desorption isotherm determination
physical parameters 700 X 700 X 800mm 40kg (excl. vacuum pump)
processing power 3 samples were measured simultaneously; 3 samples were processed
Packing list
host X1, manual X1, certificate X1, warranty card X1
After-sales instructions

Warranty days: 365 days

[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/28002/Specification.html , which is only used by users who purchase commodities from the NBCHAO.