1. Data acquisition, and can control the measurement fully automatic
2. Generate a viscous flow curve and record each measurement data
3. Support up to ten historical data comparisons
4. It can be output as an Excel format file
5. Program-controlled download runs offline
6. Can analyze data, make charts and print
Practical application of DV program-controlled offline download software
Example 1: We have a sample. When the temperature is raised to 160°C, keep the temperature for 30 minutes and collect data 10 times. Then when the temperature to 170°C, keep the temperature for 30 minutes. 30 minutes, data collection 10 times, end alarm prompt. We edited the following program following the steps above:
wtp 160 //wait for the temperature to reach 160 ℃
repeat 2
wti 1800 // constant temperature for 30 minutes
csn 10 // data collection 10 times
sti 10 //Temperature setting increased by 10 ℃
end repeat
gaa //end of alarm
repeat 2 means the program segment is repeated twice, end repeat means the end of the program segment, and gaa means the alarm prompt .
Example 2: We have a non-Newtonian liquid sample, and the rotational speed is required to start at 10 rpm, and every data acquisition is 5 times, and the speed increment is 5 rpm, that is to say, at 10 rpm, the data acquisition is 5 times, Then the speed increases by 5 rpm to 15 rpm. After collecting data for 5 times, the speed increases by 5 rpm to 20 rpm. By analogy, when the rotation speed reaches 90 rpm, after 5 data collections, the alarm prompt ends.
ssn 10 //Set the starting speed to 10RPM
repeat 16
spt 2 //At the moment when the speed changes, the data is unstable, the data of the first 2 laps is not needed, and it is idling
can 5 //collect 5 times continuously, and take the average value of 5 times as the viscosity value at the current speed
ssi 5 //Speed up by 5RPM
end repeat
gaa //end of alarm