China

NIRUN NTV-CAP1LV cone-plate viscometer Low shear rate, rotor required separately

SE
Rotor speed 0.1~ 250rpm, viscosity range 0.1~ 92000 mPa · s, Measurement accuracy +/- 1% (F. S), Plate Temperature range -2 ℃~ 120 ℃, Temperature accuracy +/- 0.1 ℃, optional rotor CPE-40, CPE-41, CPE-42, CPE-51, CPE-52.
$8591.00/set
PromotionLoading...
Coupon
Model:
Certificate

NiRun NTV-CAP1LV cone-plate viscometerSPEC

NiRun NTV-CAP1LV cone-plate viscometerDetails

NiRun NTV-CAP1LV cone-plate viscometerPacking list

SKU
NB028167
Rotor speed
0.1~250rpm
Viscosity range
0.1~92000 mPa·s
measurement Accuracy
±1%(F.S)
Plate temperature range
-2℃~120℃
Temperature Accuracy
±0.1℃
Optional rotor
CPE-40、CPE-41、CPE-42、CPE-51、CPE-52
Power Supply
AC power adapter 100V-240V
Temperature probe
PT100 temperature probe
Data Interface
Gigabit Ethernet port
NTV-CAP1LV
cone-plate viscometer
Rotor speed 0.1~ 250rpm, viscosity range 0.1~ 92000 mPa · s, Measurement accuracy +/- 1% (F. S), Plate Temperature range -2 ℃~ 120 ℃, Temperature accuracy +/- 0.1 ℃, optional rotor CPE-40, CPE-41, CPE-42, CPE-51, CPE-52.
NIRUN NTV-CAP1LV cone-plate viscometer Low shear rate, rotor required separately Picture

technical features

1.7-inch large touch screen, easy to operate, rich display information

2. Optional DV data acquisition and program control software can measure rheological properties, and the screen can directly display rheological curves

3. Display shear rate and shear stress indicators

4. Anti-static shell, automatic lifting design

5. Built-in PT100 temperature probe

6. ARM chip processor, faster data processing speed

7. Durable new bearing design

8. Powerful human-computer interaction interface

9. Gigabit network port data communication transmission, fast and reliable data transmission

10. Support external U disk data storage; single-point, continuous, and timing three storage methods are optional

11. Password protection temperature and viscosity correction coefficients, allowing users to correct temperature and viscosity correction coefficients on the premise of reliable and accurate data

12. Support infinitely variable speed measurement of viscosity, input any speed, it will directly reflect the full scale of each rotor, which is convenient for users to select test conditions

13. Automatic conversion of dynamic viscosity and kinematic viscosity

14. Arbitrary exchange of multiple viscosity units

15. A variety of user-friendly operation style interface interchange

16. Randomly comes with a powerful operation guide to help

17. Viscosity value shows continuous change

18. There is an alarm sound prompt when the measurement range is exceeded

19. Computer full-scale linear correction

20. AC power adapter 100V-240V, good anti-interference

optional accessories

CAP Viscosity Standard Solution; DV Data Acquisition and Analysis Software                          

Applications

Medium viscosity fluid: hot melt adhesive, coating, resin, starch, screen printing ink, surface treatment, organic solvent, cream, food, gel, chewing gum, plastisol, etc.

High viscosity fluids: adhesives, gels, sealants, asphalt, printing inks, chocolate, tar, molasses, pastes, synthetic polymers, epoxy polymers, vinyl esters, roofing materials

Comparison of series parameters

Rotor numberCPA-40ZCPA-41ZCPA-42ZCPA-51ZCPA-52Z
shear rate7.52.03.843.842.0
sample size0.52.01.00.50.5
NTV-CAP1LV0.1-30000.5-110000.2-60002-480003-92000
NTV-CAP1RV1-320005-1220002-6400020-51200039-983000
NTV-CAP1HA2.6-6500010-2450005-12800041-1M78-2M
NTV-CAP1HB10.5-26100039-98200020-512000163-4M314-7.8M
M=1000000; unit: shear rate rpm, sample volume ml, cp
Other parameters
Rotating speed0~250rpm


FAQ
QHello, I would like to ask what is the heating principle of this equipment, and what kind of refrigeration and heating compressor?
A No compressor, reliable semiconductor cooling and heating
QCan the Rheometer show Linear dispersion with shear force?
A Sure.
QCan this be automatically controlled?
A This one has its own temperature control system, no need for an external Thermostatic water bath
QCan the NIRUN NTV-CAP1LV cone-plate viscometer meet the SH/T0751 standard?
A Applicable, please consult the customer service of the mall for details about the SH/T0751 standard.