United States

JENCO EC330 Conductivity Meter conductivity /TDS/temperature value test

SE
Conductivity 0-20mS/cm; TDS0~10g/L
Model:

JENCO EC330 Conductivity MeterSPEC

JENCO EC330 Conductivity MeterDetails

JENCO EC330 Conductivity MeterPacking list

SKU
NB002928
Conductivity range
0.00~20.00mS/cm
EC resolution
5uS(40~2000uS), 0.05mS(2.00~15.00mS)and 0.10mS (15.00~20.00mS)
EC measurement Accuracy
±1%FS
TDS range
0.00~10.00g/L
TDS resolution
5mg/L(20~1000mg/L), 0.05g/L(1.0~7.50g/L)and 0.10g/L (7.50~10.00 g/L)
TDS measurement Accuracy
±1% F.S.
Temperature Measurement Range
0.0~99.5℃
Display resolution
0.5℃ / 1℉
Temperature Accuracy
±0.5℃ / ±1℉
Display
LCD
Calibrated point
single point
Temperature compensation range
0.0~60.0℃
Power
4 x LR-44 batteries
Battery life
More than 100 hours
shutdown mode
Automatic shutdown without operation for 10 minutes
Dimensions
193*44*25mm
Weight
Approx. 105 g
Temperature compensation method
automatic
EC330
Conductivity Meter
Conductivity 0-20mS/cm; TDS0~10g/L
JENCO EC330 Conductivity Meter conductivity /TDS/temperature value test Picture

Introduction

Ren's EC330 is an easy-to-use conductivity meter test pen. Its functions include analysis of total dissolved solids and temperature. Features include dual-screen data display, selective or automatic range selection, interchangeable conductivity meter parts and IP67 body. This product is RoHS certified.

Features

  1. Simultaneously displays data on conductivity, temperature and total dissolved solids

  2. 1 or two-point calibration

  3. Automatic temperature compensation

  4. Optional or automatic selection of Reading Ranges for conductivity and total dissolved solids

  5. Replacement conductivity meter parts

  6. 6 measurement unit options

  7. save data

Product functional structure

Appearance description

EC330 conductivity meter appearance buttons

Screen description

EC330 conductivity meter screen display

1. Low battery indicator

2. Calibration mode display

3. Measurement mode display

4. Lock mode display

5. Conductivity/TDS mode display

6. Conductivity/TDS Unit display

7. Conductivity/TDS reading

8. Temperature reading

9. Temperature Unit display 

plan

Display diagram of JENCO EC330 conductivity meter from the United States

FAQ
QIs this the Conductivity Meter mentioned in HG/T2-1047?
A OK
QCan EC330 Pen Conductivity Meter Test Industrial Wastewater?
A Hello, it's OK.
QIs the temperature compensation of the JENCO EC330 Conductivity Meter automatic? What is the temperature range?
A Hello, it is automatic temperature compensation, and the temperature compensation fee ranges from 0 to 60 ° C.
QHow to replace the electrode of EC330 Conductivity Meter?
A Hello, the EC330 Conductivity Meter replacement electrode method is as follows: 1. Unscrew the electrode head fixing nut counterclockwise. 2. Unplug the electrode head (damaged) from the electrode socket. 3. Plug in the new electrode head and ensure that the electrode head is in good contact with the whole machine. 4. Put on the electrode head fixing nut and tighten it clockwise. 5. Soak the electrode head in distillation water for 10 minutes, and then use it after re-correction.
QWhat is the test range of this EC330 Conductivity Meter? Is it automatic or manual?
A Hello, Ren's EC330 Conductivity detector range is 0.00~ 20.00mS/cm, which is automatic temperature compensation.
QWhat water quality parameters can the jenco EC330 Conductivity Meter test?
A Hello, you can test Conductivity detector rate, thermometer total solid dissolved amount data.
Standard