| Lighting method | D/8 (diffuse illumination, 8 ° direction reception); SCI/SCE measurement; including UV/exclude UV measurement; conforms to CIE No. 15, GB/T 3978, GB 2893, GB/T 18833, ISO7724-1, ASTM E1164, DIN5033 Teil7 |
| Integrating sphere size | F40mm |
| lighting source | Light spectrum LED light source, UV light source |
| Spectroscopic method | Plane grating spectroscopy |
| sensor | Silicon photodiode array (double row 40 groups) |
| Wavelength range | 400~700nm |
| wavelength spacing | 10nm |
| Light spectrum half bandwidth | 10nm |
| Reflectance Measurement Range | 0~200% |
| Measurement caliber | F5mm |
| Light-containing method | Simultaneous testing of SCI/SCE |
| Color space | CIE LAB,XYZ,Yxy,LCh,CIE LUV,s-RGB,HunterLab,βxy,DIN Lab99 Munsell(C/2) |
| Color difference formula | ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00 |
| Chromaticity indicator | WI, YI, homochromatic index MI, lightfastmess, color fastness, force, coverage, 8 degree glossiness |
| Viewing Angle | CIE 2 ° / 10 ° |
| Available Light Sources | D65,A,C,D50,F2(CWF),F7(DLF),F10(TPL5),F11(TL84),F12(TL83/U30) |
| display mode | Light spectrum graphs/data, sample chromatic values, color difference values/graphs, pass/fail results, Color simulation, Color bias |
| Measurement time | About 1.5s (about 3.2s for simultaneous test SCI/SCE) |
| Spectral reflectance repeability | MAV/SCI, standard deviation within 0.08% (400~700nm: within 0.18%) |
| Inter-station difference | ΔE * from < 0.15 |
| Measurement mode | Single measurement, average measurement (2~ 99 times) |
| Positioning method | Camera positioning, stabilizer positioning |
| Dimensions | 129*76*217mm |
| Weight | Approx. 600g |
| Battery life | Lithium battery, 6000 times in 8 hours |
| Lighting source life | 5 years more than 3 million measurements |
| Display | TFT true color 3.5inch, capacitive touch screen |
| Data Interface | USB |
| Data storage | 1000 master standards and 20,000 samples (one piece of data can include SCI/SCE at the same time) |
| Operating language | Simplified Chinese, English, Traditional Chinese |
| Operating conditions | 0~40 ℃, 0~85% RH (no condensation), altitude: less than 2000m |
| Storage environment | -20~50 ℃, 0~85% RH (no condensation) |
| Light spot size | F4mm |
| Chromatic value repeability | MAV/SCI,ΔE*ab 0.03以内 |