United States

JENCO 6308DT on-line dissolved oxygen (DO) temperature controller

SE
DO (0~40.00ppm), 4~ 20mA reversible isolated output, five sets of relay control, RS-485 interface
Model:

JENCO 6308DT Dissolved oxygen (DO) temperature controllerSPEC

JENCO 6308DT Dissolved oxygen (DO) temperature controllerDetails

JENCO 6308DT Dissolved oxygen (DO) temperature controllerPacking list

SKU
NB002956
DO range
0.00~40.00ppm
DOS resolution
0.01ppm/0.1%
Salinity compensation range
0.0~49.9ppt
Temperature compensation range
-10.0~80.0℃
Pressure compensation range
600~4000mBar
Measurement range
-10.0~120.0℃
Temperature Resolution
0.1℃
temperature sensor
Thermistor, 22K/25 ℃
current output
4~ 20mA reversible, isolated current output
Insulation Voltage/maximum load
500VDC/550Ω
Control mode
In-system ON/OFF control
Maximum load
2.5A/230VAC
key
All have sound feedback effects
Dimensions
96*96*148mm
opening size
92*92mm
Protection level
IP65
Weight
960g
Power Supply
230VAC,50/60Hz
Shop Accessories
Electrode, cable, protective cover, electrode seat, sealing box
Measurement Accuracy
±0.3℃
DO Measurement Accuracy
±0.2%FS
DO% Measurement range
0.0~400.0%
6308DT
Dissolved oxygen (DO) temperature controller
DO (0~40.00ppm), 4~ 20mA reversible isolated output, five sets of relay control, RS-485 interface
JENCO 6308DT on-line dissolved oxygen (DO) temperature controller Picture

Introduction

  • Standard 1/4 DIN waterproof body design

  • 128x64 dot matrix backlit LCD display, can display pH value, temperature value at the same time

  • Polarographic dissolved oxygen sensor is used

  • 4~20mA reversible, with isolated current output, the output range can be set by the user

  • Three sets of pH relay control/one set of temperature relay control, the control point/hysteresis width can be set by the user

  • It has the function of automatic temperature compensation

  • Power-off correction value memory function

  • RS-485 communication interface is adopted

  • The products have passed CE certification

    JENCO 6308DT In-Line Dissolved Oxygen (DO) Temperature Controller

6308DT Industrial In-Line Dissolved Oxygen (DO) Temperature Controller Optional Accessory


ModelIntroduction
electrodeOXYSENS 120Sewage and aquaculture fields
OXYFERM 120High-temperature fermentation system
cableOXYFERM 120 special cable3m/5m/10m/15m line length
Sheath tube009B1 meter tube length
Special sheath for OXYFERM 120
The electrodes are seatedThe OXYFERM 120 is dedicated to seating
Sealed box011-200-00Dial-mounted
012-200-00Wall-mounted


FAQ
QThe film of the dissolved oxygen probe of JENCO6308DT is a little loose, will it affect the test error value?
A Hello, yes. When the film is loose, wrinkled, polluted, damaged or bubbled in the electrolysis solution, it will cause reading errors.
QRen's 6308DT online dissolved oxygen temperature controller uses what type of Sensors?
A Hello, Polarographic Dissolved Oxygen Sensors are used.
QIs the 6308DT dissolved oxygen temperature controller to be zero-corrected or saturation-corrected first?
A Hello, general users do not need to perform zero correction. If you want to perform zero correction (Z), you need to perform dissolved oxygen saturation correction (S) after zero correction.
QExcuse me, the current output is required to be 15mA. Does this one meet the requirements?
A Hello, the current output of the 6308DT dissolved oxygen temperature controller is user-definable and ranges from 4 to 20mA.
QDoes the jenco 6308DT need to be corrected after power failure?
A Hello, Ren's 6308DT has the function of power-off correction value memory.
Standard