JENCO 3321 online conductivity resistivity temperature transmitter

SE
0~20.0mS/cm, two sets of relay control, backlit LCD display
Model:

JENCO 3321 Resistivity temperature transmitterSPEC

JENCO 3321 Resistivity temperature transmitterDetails

JENCO 3321 Resistivity temperature transmitterPacking list

SKU
NB002963
Measurement range
0.000uS/cm~19.99mS/cm
Resolution
0.001uS/cm
Accuracy
0.5% +/- 1 word
Types of electrode constants
0.01, 0.1, 1.0 (2-wire Conductivity Detector electrode)
reference temperature
25.0 ℃, factory setting
Relay Control Output/Maximum Load
2 sets of ON/OFF relays in the system control 5A/115VAC or 2.5A/220VAC
Dimensions
1/4DIN standard body, 96 * 48 * 110mm
opening size
92*45mm
Power Supply
100~240VAC,50/60Hz
Weight
200g
Shop Accessories
Conductivity electrode, password box
3321
Resistivity temperature transmitter
0~20.0mS/cm, two sets of relay control, backlit LCD display
JENCO 3321 online conductivity resistivity temperature transmitter Picture

JENCO 3321/3331 Online Conductivity Resistivity Temperature Transmitter Features

  • Standard 1/8DIN design correction value power-off memory function

  • With LCD backlit display

  • 4~20mA Isolated Current Output(3331)

  • Group 2 (Conductivity, Resistivity) Electrical Control(3321)

  • Automatic/manual temperature compensation function

  • CE marked

JENCO 3321 Online Conductivity Resistivity Temperature Transmitter

Optional


ModelIntroduction
Conductivity electrodes392-122Pure water ultrapure water treatment field (can be used below 80°C for a long time)
392-126Pure water ultrapure water treatment field (can be used below 80°C for a long time)
392-125Circulating water and general water treatment (can be used below 80°C for a long time)
EC621-001Pure water ultrapure water treatment field (can be used below 120°C for a long time)
EC621-01Pure water ultrapure water treatment field (can be used below 120°C for a long time)
Sealed box011-150-00Dial-mounted
012-150-00Wall-mounted


FAQ
QCan the JENCO 3321 resistivity temperature transmitter meet the HG/T2003 standard?
A Applicable, please consult the customer service of the mall for details about the HG/T2003 standard.
QHow is this 3321 resistivity temperature transmitter calibrated?
A Model 3321 uses single point correction. In measurement mode, press and hold the [UP] key and the [ENTER] key at the same time, the whole machine will enter the Conductivity detector rate/resistivity correction mode, and the "CAL" font will display and display the current measurement value of a Conductivity detector rate/resistivity. Clean the Conductivity detector electrode with deionized water and put the electrode into the appropriate standard solution. After the temperature stabilizes, press the [UP] key or the [DOWN] key to adjust the current measurement value to the standard solution value, and then press the [ENTER] key to store. At this time, the Conductivity detector electrode correction of the standard solution is completed.
Standard