PORUIKE BCT-210C split high-precision Coating Thickness Gauge F1 probe

PORUIKE split high-precision Coating Thickness Gauge BCT-210C with magnetic Induction F1 probe, thickness range 0-1250μm, resolution 0.1μm, BCT-210C CHINA Film Thickness Gauge suitable for platings, paints layer, enamel layer, aluminum tile, copper tile, phosphating layer and other thickness measurement, with continuous and single Measurement mode.

PoRuiKe BCT-210C(F1) Coating Thickness GaugeSPEC

PoRuiKe BCT-210C(F1) Coating Thickness GaugeDetails

PoRuiKe BCT-210C(F1) Coating Thickness GaugePacking list

SKU
NB029999
Measurement principle
Magnetic Induction
Measurement range
0~1250μm
margin of error
+/- [3% H + 1] (one point calibrated ); +/- [( 1~ 3) % H + 1] (two point calibrated)
Resolution
0.1μm
Minimum measurement surface
Φ7mm
Minimum radius of curvature
Convex 1.5mm
thinnest substrate
0.5mm
probe
Line probe
Calibrated Certificate
none
Calibrated mode
One point calibrated, two point calibrated
Measurement mode
Continuous Measurement mode and Single Measurement mode
Unit display
μm
statistical function
Measurement times, MAX Maximum, MIN Minimum, Average, Standard deviation
Interface
USB port
Data storage
500 Measured values
Power Supply
AA (AA) battery
system language
Chinese
Display screen
LCD liquid crystal display
Operating temperature
0-40℃
Relative Humidity
20~90%RH
Connect computer
Can connect to computer (optional)
shutdown mode
Manual/Automatic
Operating Mode
Direct mode and group mode
BCT-210C(F1)
Coating Thickness Gauge
PORUIKE BCT-210C split high-precision Coating Thickness Gauge F1 probe Picture

Functional use

Perrick Split High Precision Coating Thickness Gauge BCT-210C can be used for the thickness measurement of electroplating, paint layer, enamel layer, aluminum tile, copper tile, babbitt tile, phosphate layer, paper, and can also be used for thickness measurement of hull paint and adhesion of underwater structure. This instrument can be widely used in manufacturing, metal processing industry, chemical industry, commodity inspection and other testing fields.

Features

  • Full Chinese interface operation, more convenient;

  • This instrument adopts two thickness measurement methods: magnetic and eddy current;

  • It can be equipped with 10 kinds of probes F400, F1, F1/90°, F10, CN02, N1, CN02, N10, F5

  • Limit can be set: the measured value outside the limit can be automatically alarmed, and a batch of measured values can be analyzed by histogram;

  • LCD liquid crystal display, backlight control can be switched on and off;

  • With menu operation;

  • It has the function of communicating with the PC: the measured value and statistical value can be transmitted to the PC for further processing of the data;

  • It has an error prompt function, which can be prompted by screen display or buzzer;

  • There are two shutdown modes: manual shutdown mode and automatic shutdown mode;

  • It is designed with 2 No. 5 batteries, which is easy to replace;

  • It adopts international 9-pin socket and standard USB interface, which can communicate with the host computer through software.

Coating Thickness Gauge optional probe parameter table

Probe modelF400F1F1/90F10N1CN02
Operating Principlemagnetic inductionEddy current
Measuring range (um)0-4000-12500-100000-1250 (0-40um chrome plating on copper)10-200
Low Resolution (um)0.10.1100.11
Indication errorOne Point Calibration (um)±[3%H+1]±[3%H+10]±[3%H+1.5]±[3%H+1]
Two-point calibration (um)±[(1~3)%H+0.7]±[(1~3)%H+1]±[(1~3)%H+10]±[(1~3)%H+1.5]-
Measurement conditionsMinimum radius of curvature (mm)Convex 1Convex 1.5Straight103Straight
Diameter of the minimum area of the substrate (mm)ф3ф7ф7ф40ф5ф7
Minimum Critical Thickness (mm)0.20.50.520.3Unlimited


FAQ
QCoating thickness gauge conforms to JB/T7663.2 standard?
A Coating thickness gauge applies to JB/T7663.2, please consult customer service for detailed application definition
QHello, what is the minimum measurement area for the BCT-210C Coating thickness gauge probe test?
A Hello, the test diameter of the testable sample is not less than 7mm.
Knowledge
Standard