Specifications
Instrument level pH: 0.1 grade; conductivity: 1.0 grade
Measurement parameter pH, ORP, conductivity, dissolved oxygen, temperature, depth
pH range (0.00~14.00)pH
Minimum resolution 0.01pH
Electronic unit error of indication ±0.10pH
ORP range (-1999.9-1999.9)mV
Minimum resolution 0.1mV
Electronic unit error of indication ±20.0mV
Conductivity range 0μS/cm~200mS/cm
Minimum resolution 0.001μS/cm
Electronic unit error of indication ±1.00%FS
Dissolved oxygen range (0.00~19.99)mg/L
Minimum resolution 0.01mg/L
Electronic unit error of indication Dissolved oxygen concentration : +/- 0.30 mg/L; dissolved oxygen saturation : +/- 1 0.0%
Temperature range (-5.0~70.0)℃
Minimum resolution 0.1℃
Electronic unit error of indication ±0.5℃
depth range (0.00~10.00)m
Minimum resolution 0.01m
Electronic unit error of indication ±0.05m
Size (mm), Weight (kg) 90×235×56,0.3
Packing list
After-sales instructions

Warranty days: 365 days

[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/30282/Specification.html , which is only used by users who purchase commodities from the NBCHAO.